Comparison of the standards of absorbed dose to graphite of the IRA and the BIPM for Co y rays

Details

Serval ID
serval:BIB_E97F041771D2
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Comparison of the standards of absorbed dose to graphite of the IRA and the BIPM for Co y rays
Journal
Rapport BIPM-90/9
Author(s)
Perroche A. M., Gardel P.
Publication state
Published
Issued date
01/1990
Language
french
Abstract
A comparison between the standards of absorbed dose to graphite of the
Institut de Radiophysique Appliquee and the Bureau International des Poids
et Mesures has been performed at depths from 3 to 16 g cm-2 The ratio of
the absorbed dose rates determined by each standard is 0,997 6 at the
reference depth of 5 g cm-2