Silastic implants of the scaphoid and lunate. A long-term clinical study with a mean follow-up of 13 years.

Details

Serval ID
serval:BIB_D8CDDEF8F831
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Silastic implants of the scaphoid and lunate. A long-term clinical study with a mean follow-up of 13 years.
Journal
Journal of Hand Surgery
Author(s)
Egloff D.V., Varadi G., Narakas A., Simonetta C., Cantero C.
ISSN
0266-7681
Publication state
Published
Issued date
12/1993
Peer-reviewed
Oui
Volume
18
Number
6
Pages
687-692
Language
english
Abstract
39 implant replacement arthroplasties were reviewed with a mean follow-up of 13 years. The study includes nine cases of partial replacement of the proximal pole of the scaphoid, 16 cases of total replacement of the scaphoid and 14 cases of replacement of the lunate. In 33 cases the conventional silicone elastomer (CSE) was used and in six the high performance elastomer (HP). Subjective, clinical and radiological evaluations gave different results. Respectively the results were excellent in 41, 23 and 5%, satisfactory in 41, 49 and 39% and poor in 18, 28 and 56% of the cases. Lytic lesions in carpal and/or pericarpal bones were present in all except two cases. Study of the X-rays chronologically revealed that they increased in number and size. Six cases had their implant removed. All had histologically proven silicone synovitis. A major secondary operation has been performed or is foreseen in all cases except one. We do not recommend implant replacement arthroplasty for the scaphoid or the lunate bone.
Keywords
Adolescent, Adult, Aged, Carpal Bones/surgery, Female, Follow-Up Studies, Humans, Male, Middle Aged, Prostheses and Implants/adverse effects, Reoperation, Silicone Elastomers/adverse effects, Synovitis/etiology, Time Factors