Title
9.4-14.1 T small-animal PET-MR imaging: Feasibility analysis of LYSO APD readout via long signal lines
Journal
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Author(s)
Neves J.A., Bugalho R., Gruetter R., Magill A.W., Ortigao C., Silva J.C., Silva R., Varela J.
ISSN
0168-9002
Publication state
Published
Issued date
2013
Volume
702
Pages
98-100
Language
english
Abstract
In the present work we intend to assess the readout feasibility of Avalanche Photodiode (APD) detectors via long signal lines for the development of a combined small-animal PET-MR prototype based on the ClearPEM technology. The detection performance of a LYSO-APD module was evaluated reading out the APD charge signals to the front-end ASIC via an 80 mm length flexible flat-cable (FFC). Experimental results showed a time resolution of 5.06 ns for the detector module in double-readout mode, with a nonsignificant degradation of 8.4% with the introduction of the FFC. The energy resolution of the system was not degradated by the FFC.
Keywords
Small-animal PET, ClearPEM, UHF-MRI, Simultaneous PET-MRI