Characteristics of a New X-Ray Imaging System for Interventional Procedures: Improved Image Quality and Reduced Radiation Dose.

Details

Serval ID
serval:BIB_AF9647EA10CB
Type
Article: article from journal or magazin.
Collection
Publications
Title
Characteristics of a New X-Ray Imaging System for Interventional Procedures: Improved Image Quality and Reduced Radiation Dose.
Journal
Cardiovascular and interventional radiology
Author(s)
Schernthaner R.E., Haroun R.R., Nguyen S., Duran R., Sohn J.H., Sahu S., Chapiro J., Zhao Y., Radaelli A., van der Bom I.M., Mauti M., Hong K., Geschwind J.H., Lin M.
ISSN
1432-086X (Electronic)
ISSN-L
0174-1551
Publication state
Published
Issued date
03/2018
Peer-reviewed
Oui
Volume
41
Number
3
Pages
502-508
Language
english
Notes
Publication types: Journal Article
Publication Status: ppublish
Abstract
To compare image quality and radiation exposure between a new angiographic imaging system and the preceding generation system during uterine artery embolization (UAE).
In this retrospective, IRB-approved two-arm study, 54 patients with symptomatic uterine fibroids were treated with UAE on two different angiographic imaging systems. The new system includes optimized acquisition parameters and real-time image processing algorithms. Air kerma (AK), dose area product (DAP) and acquisition time for digital fluoroscopy (DF) and digital subtraction angiography (DSA) were recorded. Body mass index was noted as well. DF image quality was assessed objectively by image noise measurements. DSA image quality was rated by two blinded, independent readers on a four-rank scale. Statistical differences were assessed with unpaired t tests and Wilcoxon rank-sum tests.
There was no significant difference between the patients treated on the new (n = 36) and the old system (n = 18) regarding age (p = 0.10), BMI (p = 0.18), DF time (p = 0.35) and DSA time (p = 0.17). The new system significantly reduced the cumulative AK and DAP by 64 and 72%, respectively (median 0.58 Gy and 145.9 Gy*cm javax.xml.bind.JAXBElement@6b0b617f vs. 1.62 Gy and 526.8 Gy*cm javax.xml.bind.JAXBElement@1c9881bf , p < 0.01 for both). Specifically, DAP for DF and DSA decreased by 59% (75.3 vs. 181.9 Gy*cm javax.xml.bind.JAXBElement@5a14c43 , p < 0.01) and 78% (67.6 vs. 312.2 Gy*cm javax.xml.bind.JAXBElement@566a5df7 , p < 0.01), respectively. The new system achieved a significant decrease in DF image noise (p < 0.01) and a significantly better DSA image quality (p < 0.01).
The new angiographic imaging system significantly improved image quality and reduced radiation exposure during UAE procedures.

Keywords
Image quality enhancement, Interventional radiology, Radiation dosage