Transcranial duplex sonography in the detection of patent foramen ovale.

Details

Serval ID
serval:BIB_675DA26C6713
Type
Article: article from journal or magazin.
Collection
Publications
Title
Transcranial duplex sonography in the detection of patent foramen ovale.
Journal
Radiology
Author(s)
Blersch W.K., Draganski B.M., Holmer S.R., Koch H.J., Schlachetzki F., Bogdahn U., Hölscher T.
ISSN
0033-8419 (Print)
ISSN-L
0033-8419
Publication state
Published
Issued date
2002
Peer-reviewed
Oui
Volume
225
Number
3
Pages
693-699
Language
english
Notes
Publication types: Comparative Study ; Journal Article
Abstract
PURPOSE: To determine the sensitivity of contrast material-enhanced transcranial color-coded sonography (c-TCCS) compared with that of contrast-enhanced transesophageal echocardiography (c-TEE) for detection of cardiac right-to-left shunt.
MATERIALS AND METHODS: Forty consecutive patients with stroke or transient ischemic attack were admitted to the hospital and were examined by using c-TCCS and c-TEE. High-intensity transient signals (HITS) were counted for 25 seconds after the end of the Valsalva maneuver, and the numbers of HITS were classified in one of four categories (zero HITS, one to 10 HITS, >10 HITS without curtain, and curtain). A statistically significant difference was calculated with the Fisher exact test.
RESULTS: HITS were counted in 21 (52%) patients by using c-TCCS and c-TEE. With both tests, no HITS were counted in 15 (38%) patients. In two (5%) patients, no HITS were counted with c-TEE but three HITS in one patient and five HITS in the other were counted with c-TCCS. In two (5%) patients, no HITS were counted with c-TCCS, but a small patent foramen ovale (PFO) was seen at c-TEE. With c-TCCS, the sensitivity was 91% (21 of 23) and the specificity was 88% (15 of 17). In 23 patients examined with c-TCCS, 14 (61%) patients had category 1 PFO; seven (30%) patients, category 2 PFO; and two (9%) patients, category 3 PFO. Mean HITS count in patients with category 1 PFO was 4.4 and that for those with category 2 PFO was 27.6.
CONCLUSION: c-TCCS is a sensitive noninvasive method for detecting cardiac right-to-left shunt and is as sensitive as c-TEE.
Keywords
Echocardiography, Transesophageal, Female, Heart Septal Defects, Atrial/ultrasonography, Humans, Ischemic Attack, Transient/ultrasonography, Male, Middle Aged, Predictive Value of Tests, Sensitivity and Specificity, Stroke/ultrasonography, Ultrasonography, Doppler, Transcranial
Pubmed
Web of science
Create date
18/01/2013 18:59
Last modification date
20/08/2019 15:22
Usage data