Hemodynamic properties of the hemopump HP14.

Details

Serval ID
serval:BIB_1D51D60DCDD2
Type
Article: article from journal or magazin.
Collection
Publications
Institution
Title
Hemodynamic properties of the hemopump HP14.
Journal
International Journal of Artificial Organs
Author(s)
Lachat M., Jaggy C., Leskosek B., von Segesser L., Zünd G., Vogt P., Turina M.
ISSN
0391-3988 (Print)
ISSN-L
0391-3988
Publication state
Published
Issued date
1999
Peer-reviewed
Oui
Volume
22
Number
3
Pages
155-159
Language
english
Notes
Publication types: In Vitro ; Journal Article
Publication Status: ppublish
Abstract
BACKGROUND: The Hemopump HP14 is a catheter-mounted, transvalvular, left ventricular assist device intended for femoral percutaneous insertion. The pump was developed for patients with postoperative or postinterventional low cardiac output and for CABG surgery on the beating heart. Little is known about the effect of afterload and hematocrit on the pump performance.
METHODS: The influence of hematocrit and afterload on the pump flow was tested using an in vitro model filled with heparinized bovine blood. Regression analysis of the pump flow with respect to three hematocrit values (20%, 30%, 40%) and ten afterload levels (30 mmHg-120 mmHg in 10 mmHg increments) was performed for all pump speed levels (n = 7).
RESULTS: At all pump speed levels reduction of afterload and hematocrit were significant predictors for increasing pump flow (p<0.001). For hematocrit values between 40% and 20% and highest pump speed, mean pump flow at lowest afterload ranged between 2.34 and 2.53 L/min; and at highest afterload between 1.31 and 1.53 L/min. For speed level 1, afterload of 120 mmHg and hematocrit of 40% there was a maximal retrograde flow of 230+/-35 ml/min.
CONCLUSIONS: Pump performance is significantly improved by both afterload and hematocrit reduction. In the weaning phase and during the removal of the device, the pump should run at a speed level of at least three to prevent retrograde flow in the pump. Estimates for pump flow in vivo can be extrapolated from our diagrams. Our results show that the Hemopump HP14 is a valuable alternative to intra-aortic balloon counterpulsation.