In vitro evaluation of bronchodilator drug delivery by jet nebulization during pediatric mechanical ventilation.

Détails

ID Serval
serval:BIB_FF2A3C3AAFB2
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
In vitro evaluation of bronchodilator drug delivery by jet nebulization during pediatric mechanical ventilation.
Périodique
Pediatric Critical Care Medicine
Auteur⸱e⸱s
Di Paolo E.R., Pannatier A., Cotting J.
ISSN
1529-7535 (Print)
ISSN-L
1529-7535
Statut éditorial
Publié
Date de publication
07/2005
Peer-reviewed
Oui
Volume
6
Numéro
4
Pages
462-469
Langue
anglais
Notes
Publication types: Comparative Study ; Evaluation Studies ; In Vitro ; Journal Article
Publication Status: ppublish
Résumé
OBJECTIVE: To determine the influence of jet nebulizer brands and nebulization mode on albuterol delivery in a mechanically ventilated pediatric lung model.
DESIGN: In vitro, laboratory study.
SETTING: Research laboratory of a university hospital.
INTERVENTIONS: Using albuterol as a marker, six jet nebulizers (Microneb NA420, Sidestream, Acorn II, Cirrus, Upmist, Micro Mist) were tested in four nebulization modes in a bench model mimicking the ventilatory pattern of a 10-kg infant (Galileo ventilator, Hamilton Medical). The amounts of albuterol deposited on the inspiratory filters at the end of the endotracheal tube were determined, as well as the pressure, flow profiles, and particle size distribution of the jet nebulizers.
MEASUREMENTS AND MAIN RESULTS: Pooling the data of the six jet nebulizer brands (n = 30) indicated that intermittent nebulization during the expiratory phase was more efficient (6.5 +/- 2.5% of the initial dose, p < .001) than intermittent nebulization during the inspiratory phase (1.9 +/- 1.2%) and continuous nebulization with air from the ventilator (4.0 +/- 1.5%) or from an external source (4.2 +/- 1.4%). The particle size distribution at 6 L x min(-1) was between 2.81 and 3.30 microm.
CONCLUSIONS: In our in vitro pediatric lung model, the quantity of inhaled drug was low. Jet nebulizer brands and nebulization modes significantly affected drug delivery, and in vitro models designed for adults cannot be extrapolated to infants.
Mots-clé
Aerosols, Albuterol/administration & dosage, Bronchodilator Agents/administration & dosage, Humans, Infant, Nebulizers and Vaporizers, Particle Size, Respiration, Artificial, Technology Assessment, Biomedical
Pubmed
Création de la notice
25/01/2008 10:58
Dernière modification de la notice
20/08/2019 17:29
Données d'usage