Improved phosphorus MRSI acquisition through compressed sensing acceleration combined with low-rank reconstruction.

Détails

ID Serval
serval:BIB_FADBE8F854E9
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Improved phosphorus MRSI acquisition through compressed sensing acceleration combined with low-rank reconstruction.
Périodique
Magma
Auteur⸱e⸱s
Songeon J., Lazeyras F., Agius T., Dabrowski O., Ruttimann R., Toso C., Longchamp A., Klauser A., Courvoisier S.
ISSN
1352-8661 (Electronic)
ISSN-L
0968-5243
Statut éditorial
Publié
Date de publication
04/2025
Peer-reviewed
Oui
Volume
38
Numéro
2
Pages
161-173
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
Phosphorus-31 magnetic resonance spectroscopic imaging ( <sup>31</sup> P-MRSI) is a non-invasive tool for assessing cellular high-energy metabolism in-vivo. However, its acquisition suffers from a low sensitivity, which necessitates large voxel sizes or multiple averages to achieve an acceptable signal-to-noise ratio (SNR), resulting in long scan times.
To overcome these limitations, we propose an acquisition and reconstruction scheme for FID-MRSI sequences. Specifically, we employed Compressed Sensing (CS) and Low-Rank (LR) with Total Generalized Variation (TGV) regularization in a combined CS-LR framework. Additionally, we used a novel approach to k-space undersampling that utilizes distinct pseudo-random patterns for each average. To evaluate the proposed method's performance, we performed a retrospective analysis on healthy volunteers' brains and ex-vivo perfused kidneys.
The presented method effectively improves the SNR two-to-threefold while preserving spectral and spatial quality even with threefold acceleration. We were able to recover signal attenuation of anatomical information, and the SNR improvement was obtained while maintaining the metabolites peaks linewidth.
We presented a novel combined CS-LR acceleration and reconstruction method for FID-MRSI sequences, utilizing a unique approach to k-space undersampling. Our proposed method has demonstrated promising results in enhancing the SNR making it applicable for reducing acquisition time.
Mots-clé
Humans, Signal-To-Noise Ratio, Phosphorus/chemistry, Brain/diagnostic imaging, Brain/metabolism, Image Processing, Computer-Assisted/methods, Algorithms, Magnetic Resonance Spectroscopy/methods, Kidney/diagnostic imaging, Kidney/metabolism, Magnetic Resonance Imaging/methods, Retrospective Studies, Data Compression/methods, Healthy Volunteers, Phantoms, Imaging, Adult, Male, Acceleration, Compressed sensing, Low rank, Phosphorus magnetic resonance spectroscopic imaging (31P-MRSI), Regularization