Comparison of methods for detection of plasmid-mediated and chromosomally encoded colistin resistance in Enterobacteriaceae.

Détails

Ressource 1Télécharger: 28606644AM.pdf (542.12 [Ko])
Etat: Public
Version: Author's accepted manuscript
ID Serval
serval:BIB_F44C247A767D
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Comparison of methods for detection of plasmid-mediated and chromosomally encoded colistin resistance in Enterobacteriaceae.
Périodique
Clinical microbiology and infection
Auteur⸱e⸱s
Jayol A., Nordmann P., Lehours P., Poirel L., Dubois V.
ISSN
1469-0691 (Electronic)
ISSN-L
1198-743X
Statut éditorial
Publié
Date de publication
02/2018
Peer-reviewed
Oui
Volume
24
Numéro
2
Pages
175-179
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
Because of the emergence of plasmid-mediated (mcr-1 and mcr-2 genes) and chromosomally encoded colistin resistance, reliable methods for detecting colistin resistance/susceptibility in routine laboratories are required. We evaluated the respective performances of the BD Phoenix automated system, the newly developed Rapid Polymyxin NP test and the broth microdilution (BMD) reference method to detect colistin resistance in Enterobacteriaceae, and particularly those producing mcr-1 and mcr-2.
Colistin susceptibility of 123 enterobacterial clinical isolates (40 colistin-susceptible and 83 colistin-resistant isolates) was tested with the BD Phoenix automated system, the Rapid Polymyxin NP test and the BMD method. Molecular mechanisms responsible for plasmid-mediated and chromosomally encoded colistin resistance mechanisms were investigated by PCR and sequencing.
Considering BMD as a reference method, the BD Phoenix system failed to detect ten colistin-resistant isolates (one Escherichia coli, one Klebsiella pneumoniae, seven Enterobacter species and one Salmonella enterica). The Rapid Polymyxin NP test failed to detect the same single E. coli isolate. Those two latter methods detected the 16 E. coli, K. pneumoniae and S. enterica isolates producing the plasmid-encoded mcr-1 and mcr-2.
The BD Phoenix system and the Rapid Polymyxin NP test are reliable techniques for detecting plasmid-mediated mcr-1 and mcr-2-related colistin resistance. However, a high rate of false susceptibility was observed with the BD Phoenix system, indicating that susceptibility results obtained with that system should be confirmed by BMD method. By contrast, the Rapid Polymyxin NP test showed a good agreement with the BMD method, and results were obtained rapidly (within 2 hours). The BMD method should be performed if minimum inhibitory concentration values are needed.

Mots-clé
BD Phoenix automated system, Colistin, Polymyxin B, Rapid Polymyxin NP test, Resistance, Susceptibility testing, mcr-1
Pubmed
Web of science
Création de la notice