One-tissue compartment model for myocardial perfusion quantification with N-13 ammonia PET provides matching results: A cross-comparison between Carimas, FlowQuant, and PMOD.

Détails

Ressource 1Télécharger: 34409572_BIB_F26CB8E83CA4.pdf (584.74 [Ko])
Etat: Public
Version: Final published version
Licence: CC BY 4.0
ID Serval
serval:BIB_F26CB8E83CA4
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
One-tissue compartment model for myocardial perfusion quantification with N-13 ammonia PET provides matching results: A cross-comparison between Carimas, FlowQuant, and PMOD.
Périodique
Journal of nuclear cardiology
Auteur⸱e⸱s
Nesterov S.V., Sciagrà R., Orozco LEJ, Prior J.O., Settimo L., Han C., Deshayes E., deKemp R.A., Ryzhkova D.V., Gwet K.L., Knuuti J.M.
ISSN
1532-6551 (Electronic)
ISSN-L
1071-3581
Statut éditorial
Publié
Date de publication
10/2022
Peer-reviewed
Oui
Volume
29
Numéro
5
Pages
2543-2550
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
To cross-compare three software packages (SPs)-Carimas, FlowQuant, and PMOD-to quantify myocardial perfusion at global, regional, and segmental levels.
Stress N-13 ammonia PET scans of 48 patients with HCM were analyzed in three centers using Carimas, FlowQuant, and PMOD. Values agreed if they had an ICC > 0.75 and a difference < 20% of the median across all observers.
When using 1TCM on the global level, the agreement was good, and the maximum difference between 1TCM MBF values was 17.2% (ICC = 0.83). On the regional level, the agreement was acceptable except in the LCx region (25.5% difference, ICC = 0.74) between FlowQuant and PMOD. Carimas-1TCM agreed well with PMOD-1TCM and FlowQuant-1TCM. Values obtained with FlowQuant-1TCM had a somewhat lesser agreement with PMOD-1TCM, especially at the segmental level.
The global and regional MBF values (with one exception) agree well between the different software packages. There is significant variability in segmental values, mainly located in the LCx region and segments. Out of the studied tools, Carimas can be used interchangeably with both PMOD and FlowQuant for 1TCM implementation on all levels-global, regional, and segmental.
Mots-clé
Ammonia, Coronary Circulation, Humans, Myocardial Perfusion Imaging, Perfusion, Positron-Emission Tomography, Reproducibility of Results, Software, Imaging software, Myocardial perfusion imaging, N-13 ammonia, absolute quantification, agreement, positron emission tomography, quantitative imaging, reproducibility, standardization of PET