Discrepancy between radioimmunoassay and high performance liquid chromatography tandem-mass spectrometry for the analysis of androstenedione.

Détails

ID Serval
serval:BIB_ED48DC770D39
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Titre
Discrepancy between radioimmunoassay and high performance liquid chromatography tandem-mass spectrometry for the analysis of androstenedione.
Périodique
Analytical Biochemistry
Auteur⸱e⸱s
Bruce S.J., Rey F., Béguin A., Berthod C., Werner D., Henry H.
ISSN
1096-0309 (Electronic)
ISSN-L
0003-2697
Statut éditorial
Publié
Date de publication
2014
Peer-reviewed
Oui
Volume
455
Pages
20-25
Langue
anglais
Notes
Publication types: Journal Article Publication Status: ppublish
Résumé
The discrepancy of results for the quantification of androstenedione in human serum between a radioimmunoassay (RIA) method and high performance liquid chromatography tandem-mass spectrometry (LC-MS/MS) was investigated. RIA overestimated concentrations compared to LC-MS/MS on 59 clinical samples (RIA=1.79×LC-MS/MS+0.94). RIA kit and LC-MS/MS calibrants were also determined by both methods. The RIA performed with improved accuracy on the calibrants (RIA=1.35×LC-MS/MS-0.28). Lipid, protein, electrolyte content, and pH of the two sets of calibrants were further investigated. The RIA calibrants contained little lipid material, while the LC-MS/MS calibrant material contained the same levels expected in normal serum/plasma. The pH and sex hormone binding globulin (SHBG) values were different between the RIA calibrants and the LC-MS/MS calibrant material (SHBG, 31±2 and 38±2nmol/l; pH, 8.27±0.18 and 8.66±0.03, respectively). No correlation was observed between androstenedione RIA and LC-MS/MS discrepancy and lipid or protein. LC-MS/MS sample preparation was tested for the removal of protein-bound material and recovery determined (99-108%). The corresponding RIA results overestimated androstenedione by 52-174% compared to LC-MS/MS. The results here demonstrate that LC-MS/MS is the more accurate method.
Pubmed
Web of science
Création de la notice
14/06/2014 15:06
Dernière modification de la notice
20/08/2019 17:15
Données d'usage