Résumé
A retrospective radiological study.
The aim of this study was to evaluate the accuracy of pedicle screw insertion using O-Arm navigation, robotic assistance, or a freehand fluoroscopic technique.
Pedicle screw insertion using either "O-Arm" navigation or robotic devices is gaining popularity. Although several studies are available evaluating each of those techniques separately, no direct comparison has been attempted.
Eighty-four patients undergoing implantation of 569 lumbar and thoracic screws were divided into three groups. Eleven patients (64 screws) had screws inserted using robotic assistance, 25 patients (191 screws) using the O-arm, while 48 patients (314 screws) had screws inserted using lateral fluoroscopy in a freehand technique. A single experienced spine surgeon assisted by a spinal fellow performed all procedures. Screw placement accuracy was assessed by two independent observers on postoperative computed tomography (CTs) according to the A to D Rampersaud criteria.
No statistically significant difference was noted between the three groups. About 70.4% of screws in the freehand group, 69.6% in the O arm group, and 78.8% in the robotic group were placed completely within the pedicle margins (grade A) (P > 0.05). About 6.4% of screws were considered misplaced (grades C&D) in the freehand group, 4.2% in the O-arm group, and 4.7% in the robotic group (P > 0.05). The spinal fellow inserted screws with the same accuracy as the senior surgeon (P > 0.05).
The advent of new technologies does not appear to alter accuracy of screw placement in our setting. Under supervision, spinal fellows might perform equally well to experienced surgeons using new tools. The lack of difference in accuracy does not imply that the above-mentioned techniques have no added advantages. Other issues, such as surgeon/patient radiation, fiddle factor, teaching suitability, etc., outside the scope of our present study, need further assessment.