Use of active personal dosemeters in interventional radiology and cardiology: Tests in laboratory conditions and recommendations - ORAMED project

Détails

ID Serval
serval:BIB_DBD069A544A0
Type
Actes de conférence (partie): contribution originale à la littérature scientifique, publiée à l'occasion de conférences scientifiques, dans un ouvrage de compte-rendu (proceedings), ou dans l'édition spéciale d'un journal reconnu (conference proceedings).
Collection
Publications
Institution
Titre
Use of active personal dosemeters in interventional radiology and cardiology: Tests in laboratory conditions and recommendations - ORAMED project
Titre de la conférence
ORAMED 2011, International Workshop on Optimization of Radiation Protection of Medical Staff
Auteur⸱e⸱s
Clairand I., Bordy J.-M., Carinou E., Daures J., Debroas J., Denozière M., Donadille L., Ginjaume M., Itié C., Koukorava C., Krim S., Lebacq A.-L., Martin P., Struelens L., Sans-Merce M., Vanhavere F.
Adresse
Barcelona, Spain, January 20-22, 2011
ISBN
1350-4487
Statut éditorial
Publié
Date de publication
2011
Peer-reviewed
Oui
Volume
46
Série
Radiation Measurements
Pages
1252-1257
Langue
anglais
Résumé
Active personal dosemeters (APD) have been found to be very efficient tools to reduce occupational doses in many applications of ionizing radiation. In order to be used in interventional radiology and cardiology (IR/IC), APDs should be able to measure low energy photons and pulsed radiation with relatively high instantaneous personal dose equivalent rates. A study concerning the optimization of the use of APDs in IR/IC was performed in the framework of the ORAMED project, a Collaborative Project (2008-2011) supported by the European Commission within its 7th Framework Program. In particular, eight commercial APDs were tested in continuous and pulsed X-ray fields delivered by calibration laboratories in order to evaluate their performances. Most of APDs provide a response in pulsed mode more or less affected by the personal dose equivalent rate, which means they could be used in routine monitoring provided that correction factors are introduced. These results emphasize the importance of adding tests in pulsed mode in type-test procedures for APDs. Some general recommendations are proposed in the end of this paper for the selection and use of APDs at IR/IC workplaces.