Intercomparison of activity measurements for beta-emitters in Swiss nuclear medicine laboratories.

Détails

ID Serval
serval:BIB_DBC46990849E
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Intercomparison of activity measurements for beta-emitters in Swiss nuclear medicine laboratories.
Périodique
Journal of Nuclear Medicine Technology
Auteur⸱e⸱s
Wastiel C., Valley J.F., Delaloye A.B., Leresche M., Linder R., Sassowsky M., Bochud F.O.
ISSN
0091-4916 (Print)
ISSN-L
0091-4916
Statut éditorial
Publié
Date de publication
2005
Volume
33
Numéro
4
Pages
238-242
Langue
anglais
Notes
Publication types: Comparative Study ; Journal Article ; Multicenter Study
Publication Status: ppublish
Résumé
OBJECTIVE: The activity of injected radiopharmaceuticals in nuclear medicine, including beta-emitters used for pain palliation, has to be monitored systematically. The objective of the present work was to evaluate the situation and precision of activity monitoring for beta-emitters in Swiss nuclear medicine laboratories.
METHODS: A questionnaire about the monitoring methods used was sent to 50 centers. On the basis of the questionnaire results, an intercomparison of activity measurements with 90Y and 169Er sources was organized.
RESULTS: This study showed that most laboratories check beta-emitter activity with a dose calibrator measurement in the original vial provided by the producer or in the injection syringe. They therefore need to have calibration factors for the corresponding measurement geometries. The results of the intercomparison were disappointing overall. Sixteen of 27 90Y measurements and 17 of 22 169Er measurements in the original vial deviated from the reference activity by more than 20%. The situation was similar for the syringe. These discrepancies did not stem from the intrinsic limitation of the measuring method but were mainly attributable to the poor quality of the calibration factors provided by the manufacturers, in addition to lack of follow-up and incorrect background subtraction, particularly for 169Er, by the nuclear medicine laboratories. Manufacturers are being contacted to discuss possible improvements for the situation.
CONCLUSION: This study showed that commercial dose calibrators are generally adequate for measurement of the activities of beta-emitters. However, in some cases, the measurement of 90Y can lead to errors reaching +/-50%. For 169Er, with its much lower beta-energy, the situation is even worse; the observed differences can be higher than 1 order of magnitude.
Mots-clé
Beta Particles, Calibration/standards, Laboratories/standards, Laboratories/statistics & numerical data, Nuclear Medicine/standards, Nuclear Medicine/statistics & numerical data, Quality Assurance, Health Care/standards, Quality Assurance, Health Care/statistics & numerical data, Questionnaires, Radiation Dosage, Radiometry/methods, Radiometry/standards, Radiopharmaceuticals/analysis, Radiopharmaceuticals/standards, Reference Values, Switzerland/epidemiology
Pubmed
Création de la notice
25/04/2008 17:11
Dernière modification de la notice
20/08/2019 16:00
Données d'usage