Assessment of the carotid artery by MRI at 3T: a study on reproducibility
Détails
ID Serval
serval:BIB_D55886964D8D
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Assessment of the carotid artery by MRI at 3T: a study on reproducibility
Périodique
Journal of Magnetic Resonance Imaging
ISSN
1053-1807
Statut éditorial
Publié
Date de publication
2007
Peer-reviewed
Oui
Volume
25
Numéro
5
Pages
1035-1043
Langue
anglais
Résumé
PURPOSE: To examine the reproducibility of carotid artery dimension measurements using 3T MRI. MATERIALS AND METHODS: Ten healthy volunteers underwent three scans on two occasions for assessment of total vessel wall area (TVWA), total luminal area (TLA), and minimum (MinT) and maximum (MaxT) vessel wall thickness. A double inversion-recovery (IR) fast gradient-echo (FGRE) sequence was used on a commercial 3T system. During the first visit the subjects were scanned twice. The third scan was performed at least four days later. One observer traced all scans, and a second observer retraced the first scan series. RESULTS: For TVWA an interclass correlation (ICC) of 0.994 was calculated with all three scans taken into account. The interobserver ICC was 0.984. The agreement between the scans for TLA showed an ICC of 0.982 with an interobserver ICC of 0.998. For MinT and MaxT an ICC of 0.843 and 0.935 were calculated, with interobserver ICCs of 0.860 and 0.726, respectively. CONCLUSION: With the use of a commercial 3T MR system, TVWA, TLA, and wall thickness measurements of the carotid artery can be assessed with good reproducibility.
Mots-clé
Adult, Aged, Carotid Arteries/*anatomy & histology, Female, Humans, Image Processing, Computer-Assisted, Magnetic Resonance Angiography/*methods, Male, Middle Aged, Reproducibility of Results
Pubmed
Web of science
Open Access
Oui
Création de la notice
02/03/2010 16:04
Dernière modification de la notice
20/08/2019 15:55