Impact of the bicuspid aortic valve on aortic root haemodynamics: three-dimensional computed fluid dynamics simulation.

Détails

ID Serval
serval:BIB_CE3308AFE53E
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Impact of the bicuspid aortic valve on aortic root haemodynamics: three-dimensional computed fluid dynamics simulation.
Périodique
Interactive cardiovascular and thoracic surgery
Auteur⸱e⸱s
Berdajs D., Mosbahi S., Eckstein F.S., Charbonnier D., Ferrari E., von Segesser L.K.
ISSN
1569-9285 (Electronic)
ISSN-L
1569-9285
Statut éditorial
Publié
Date de publication
01/09/2018
Peer-reviewed
Oui
Volume
27
Numéro
3
Pages
446-454
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
The aim was to evaluate the impact of a bicuspid aortic valve (BAV) on local shear stress and on the pressure profile on the elements of the aortic root (AoR).
The experiment setup included a BAV with aortic valve stenosis (n = 5 pigs, 67 ± 3.5 kg) and insufficiency (n = 5 pigs, 66.7 ± 4.4 kg). By implanting 6 high-fidelity microsonometric crystals in each AoR, we determined the 3-dimensional (3D) geometry of the AoR. Experimental and geometric data were used to create a 3D time- and pressure-dependent computed fluid dynamic model of the AoR with the BAV.
3D AoR geometry was determined by AoR tilt (α) and rotation angle (β). Both values were maximal at the end of diastole: 24.41 ± 1.70° (α) and 20.90 ± 2.11° (β) for BAV with stenosis and 31.92 ± 11.51° (α) and 20.84 ± 9.80° (β) for BAV with insufficiency and minimal at peak ejection 23.42 ± 1.65° (α), 20.38 ± 1.61° (β) for stenosis and 26.62 ± 7.86° (α), 19.79 ± 8.45° (β) for insufficiency. In insufficiency, low shear stress of 0-0.08 Pa and moderate pressure (60-80 mmHg) were present. In BAV with stenosis, low shear stress of 0-0.5 Pa and moderate pressure (0-20 mmHg) were present at diastole; at peak ejection high shear stress >2 Pa and elevated pressure of >80 mmHg were present.
In a BAV with aortic valve stenosis, the haemodynamics are less favourable. The elevated pressure with elevated shear stress may over the long term promote degenerative processes in the leaflets and consequently valve function failure.
Mots-clé
Animals, Aorta/physiopathology, Aortic Valve/abnormalities, Aortic Valve/physiopathology, Aortic Valve Stenosis/complications, Aortic Valve Stenosis/physiopathology, Disease Models, Animal, Heart Valve Diseases/complications, Heart Valve Diseases/physiopathology, Hemodynamics, Hydrodynamics, Stress, Mechanical, Swine
Pubmed
Web of science
Open Access
Oui
Création de la notice
29/03/2018 18:43
Dernière modification de la notice
20/08/2019 16:48
Données d'usage