Effects of different levels of CT iterative reconstruction on low-contrast detectability and radiation dose in patients of different sizes: an anthropomorphic phantom study.

Détails

ID Serval
serval:BIB_C9A7903C464D
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Effects of different levels of CT iterative reconstruction on low-contrast detectability and radiation dose in patients of different sizes: an anthropomorphic phantom study.
Périodique
La Radiologia medica
Auteur⸱e⸱s
Rampado O., Depaoli A., Marchisio F., Gatti M., Racine D., Ruggeri V., Ruggirello I., Darvizeh F., Fonio P., Ropolo R.
ISSN
1826-6983 (Electronic)
ISSN-L
0033-8362
Statut éditorial
Publié
Date de publication
01/2021
Peer-reviewed
Oui
Volume
126
Numéro
1
Pages
55-62
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
The purpose of this study was to verify the maintenance of low-contrast detectability at different CT dose reduction levels, in patients of different sizes, as a consequence of the application of iterative reconstruction at different strengths combined with tube current modulation.
Anthropomorphic abdominal phantoms of two sizes (small and large) were imaged at a fixed noise with iterative algorithm ASIR-V percentages in the range between 0 and 70% and corresponding dose reductions in the range of 0-83%. A total of 1400 images with and without liver low-contrast simulated lesions were evaluated by five radiologists, using the receiver operating characteristics (ROC) paradigm and evaluating the area under the ROC curve (AUC). The human observer results were then compared with AUC obtained with a channelized Hotelling observer (CHO). CNR values were also calculated.
For the small phantom, the AUC values lie between 0.90 and 0.93 for human evaluations of images acquired without iterative reconstruction, with 30% ASIR-V and with 50% ASIR-V. The AUC decreased significantly to 0.81 (p = 0.0001) at 70% ASIR-V. The CHO results were in coherence with human observer scores. Also, similar results were observed for the large size phantom. CNR values were stable for the different ASIR-V percentages.
The iterative algorithm maintained the low-contrast detectability up to a dose reduction of about 70%, following application of a 50% ASIR-V combined with automatic tube current modulation, regardless of the phantom size. At further dose reductions using greater iterative percentages, a significant decrease in detectability was observed.
Mots-clé
Abdomen/diagnostic imaging, Algorithms, Anthropometry, Body Size, Humans, Phantoms, Imaging, Radiation Dosage, Radiographic Image Interpretation, Computer-Assisted/methods, Tomography, X-Ray Computed, Computed tomography, Iterative reconstruction, Low-contrast detectability, Model observer, Patient dose reduction
Pubmed
Web of science
Création de la notice
10/06/2020 19:51
Dernière modification de la notice
18/03/2022 6:34
Données d'usage