Ex vivo testing of the Quart arterial line filter

Détails

ID Serval
serval:BIB_C86F6C6429A5
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Ex vivo testing of the Quart arterial line filter
Périodique
Perfusion
Auteur⸱e⸱s
Mueller  X. M., Tevaearai  H. T., Jegger  D., Augstburger  M., Burki  M., von Segesser  L. K.
ISSN
0267-6591 (Print)
Statut éditorial
Publié
Date de publication
11/1999
Volume
14
Numéro
6
Pages
481-7
Notes
Journal Article --- Old month value: Nov
Résumé
Arterial line filters are now routinely used in cardiac surgery in order to decrease the microemboli load to the patient. The Quart filter (Jostra, Hirrlingen, Germany) with a new planar construction design, an easy de-airing system and an integrated bypass, was tested for air filtration capacity and resistance to blood path in a standardized setting with surviving animals. Three calves (mean body weight: 71+/-3.4 kg) were connected to a standard cardiopulmonary bypass (CPB) circuit by jugular venous and carotid arterial cannulation with a mean flow rate of 3.5 l/min. The arterial line filter was challenged with upstream injections of boluses of air of 5, 10 and 15 ml, respectively. A Doppler ultrasound was positioned downstream on the arterial line to measure bubble count and size. The pressure drop through the filter was monitored at flow rates of between 1 and 6 l/min. At the end of the procedure the animals were weaned from the CPB and, thereafter, from the ventilator. After 7 days, the animals were sacrificed electively. This study shows that important quantities of air can be injected into the arterial line upstream of the filter with small volumes of small sized bubbles recorded downstream. With the 5 ml air bolus injection, mean values of 0.3+/-0.6 bubbles of 30 and 40 microm were detected, whereas with the 20 ml bolus, 32.6+/-8.7 bubbles of 10 microm, 3.7+/-1.1 bubbles of 30 microm, 3.3+/-0.6 bubbles of 40 microm and 0.7+/-1.1 bubbles of 50 microm were recorded. The blood path resistance at different blood flow rates was well within the acceptable range with a pressure drop of 20+/-0 and 26.6+/-5.7 mmHg at flow rates of 4 and 5 l/min, respectively. With its planar concept, the Quart filter offers good air filtering capacity both in terms of bubble count and size after injection of large boluses of air, without any increase of resistance to the blood path. Moreover, it offers a venting function and an integrated bypass system.
Mots-clé
Air Animals Cardiopulmonary Bypass Catheterization Catheters, Indwelling Cattle Equipment Design Equipment Safety Filtration Heart Hemofiltration/*instrumentation Kidney Liver Lung Perfusion Pressure Spleen
Pubmed
Web of science
Création de la notice
28/01/2008 10:28
Dernière modification de la notice
20/08/2019 16:43
Données d'usage