Imposed Work of Breathing During High-Frequency Oscillatory Ventilation in Spontaneously Breathing Neonatal and Pediatric Models.
Détails
ID Serval
serval:BIB_B916484AEBF5
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Imposed Work of Breathing During High-Frequency Oscillatory Ventilation in Spontaneously Breathing Neonatal and Pediatric Models.
Périodique
Respiratory care
ISSN
1943-3654 (Electronic)
ISSN-L
0020-1324
Statut éditorial
Publié
Date de publication
09/2018
Peer-reviewed
Oui
Volume
63
Numéro
9
Pages
1085-1093
Langue
anglais
Notes
Publication types: Comparative Study ; Journal Article
Publication Status: ppublish
Publication Status: ppublish
Résumé
High-frequency oscillatory ventilation (HFOV) is used in cases of neonatal and pediatric acute respiratory failure, sometimes even as the primary ventilatory mode. Allowing patients (at least neonates) on HFOV to breathe spontaneously soon after intubation has been shown to be feasible, and this is becoming a more generally used approach for infants and small children. However, such an approach may increase the imposed work of breathing (WOB), raising the question of whether the imposed WOB varies with the use of newer-generation HFOV devices, which operate according to different functional principles.
A bench test was designed to compare the pressure-time product (PTP), a surrogate marker of the imposed WOB, produced with the use of 7 HFOV devices. Scenarios corresponding to various age groups (preterm newborn [1 kg], term newborn [3.5 kg], infant [10 kg], and child [25 kg]) and 2 respiratory system conditions (physiologic and pathologic) were tested.
The PTP varied between devices and increased with the oscillation frequency for all devices, independent of the respiratory system condition. Furthermore, the PTP increased with age and was higher for physiologic than for pathologic respiratory system conditions. We considered a change of ≥ 20% as being of clinically relevant; the effect of oscillation frequency was the most important parameter influencing imposed WOB during spontaneous breathing.
Variations in imposed WOB, as expressed by PTP values, during spontaneous breathing depend mainly on the oscillator frequency, respiratory system condition, and, though to a lesser extent, on the device itself.
A bench test was designed to compare the pressure-time product (PTP), a surrogate marker of the imposed WOB, produced with the use of 7 HFOV devices. Scenarios corresponding to various age groups (preterm newborn [1 kg], term newborn [3.5 kg], infant [10 kg], and child [25 kg]) and 2 respiratory system conditions (physiologic and pathologic) were tested.
The PTP varied between devices and increased with the oscillation frequency for all devices, independent of the respiratory system condition. Furthermore, the PTP increased with age and was higher for physiologic than for pathologic respiratory system conditions. We considered a change of ≥ 20% as being of clinically relevant; the effect of oscillation frequency was the most important parameter influencing imposed WOB during spontaneous breathing.
Variations in imposed WOB, as expressed by PTP values, during spontaneous breathing depend mainly on the oscillator frequency, respiratory system condition, and, though to a lesser extent, on the device itself.
Mots-clé
Child, Child, Preschool, Female, High-Frequency Ventilation/instrumentation, Humans, Infant, Infant, Newborn, Infant, Premature, Lung/physiopathology, Male, Models, Anatomic, Pressure, Respiration, Respiratory Function Tests, Time Factors, Work of Breathing, high-frequency oscillatory ventilation, neonatal intensive care, respiratory insufficiency, spontaneous breathing, work of breathing
Pubmed
Web of science
Création de la notice
31/07/2018 14:38
Dernière modification de la notice
20/08/2019 15:27