Synchronization device for electrocardiography-gated echo-planar imaging.

Détails

ID Serval
serval:BIB_B502EB305A70
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Titre
Synchronization device for electrocardiography-gated echo-planar imaging.
Périodique
Radiology
Auteur⸱e⸱s
Felblinger J., Debatin J.F., Boesch C., Gruetter R., McKinnon G.C.
ISSN
0033-8419 (Print)
ISSN-L
0033-8419
Statut éditorial
Publié
Date de publication
1995
Volume
197
Numéro
1
Pages
311-313
Langue
anglais
Notes
Publication types: Journal ArticlePublication Status: ppublish. PDF type: Note
Résumé
An electrocardiography (ECG) synchronization technique allowed triggering of 1.5-T echo-planar acquisitions of the heart, with high gradient slew rates. In 51 volunteers (37 men and 14 women, aged 21-48 years), the ECG signal was amplified, filtered, and converted into an optical signal directly above the heart and was transmitted optically outside the bore. Reliable and artifact-free ECG tracings were obtained in all cases, regardless of the gradient switching speed.
Mots-clé
Adult, Electrocardiography/instrumentation, Female, Humans, Magnetic Resonance Imaging/instrumentation, Male, Middle Aged
Pubmed
Web of science
Création de la notice
04/08/2010 16:28
Dernière modification de la notice
20/08/2019 16:23
Données d'usage