High dose-per-pulse electron beam dosimetry: Commissioning of the Oriatron eRT6 prototype linear accelerator for preclinical use.

Détails

ID Serval
serval:BIB_AECCCBDB6D56
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
High dose-per-pulse electron beam dosimetry: Commissioning of the Oriatron eRT6 prototype linear accelerator for preclinical use.
Périodique
Medical physics
Auteur⸱e⸱s
Jaccard M., Durán M.T., Petersson K., Germond J.F., Liger P., Vozenin M.C., Bourhis J., Bochud F., Bailat C.
ISSN
2473-4209 (Electronic)
ISSN-L
0094-2405
Statut éditorial
Publié
Date de publication
02/2018
Peer-reviewed
Oui
Volume
45
Numéro
2
Pages
863-874
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
The Oriatron eRT6 is an experimental high dose-per-pulse linear accelerator (linac) which was designed to deliver an electron beam with variable dose-rates, ranging from a few Gy/min up to hundreds of Gy/s. It was built to study the radiobiological effects of high dose-per-pulse/dose-rate electron beam irradiation, in the context of preclinical and cognitive studies. In this work, we report on the commissioning and beam monitoring of the Oriatron eRT6 prototype linac.
The beam was characterized in different steps. The output stability was studied by performing repeated measurements over a period of 20 months. The relative output variations caused by changing beam parameters, such as the temporal electron pulse width, the pulse repetition frequency and the pulse amplitude were also analyzed. Finally, depth dose curves and field sizes were measured for two different beam settings, resulting in one beam with a conventional radiotherapy dose-rate and one with a much higher dose-rate. Measurements were performed with Gafchromic EBT3 films and with a PTW Advanced Markus ionization chamber. In addition, we developed a beam current monitoring system based on the signals from an induction torus positioned at the beam exit of the waveguide and from a graphite beam collimator.
The stability of the output over repeated measurements was found to be good, with a standard deviation smaller than 1%. However, non-negligible day-to-day variations of the beam output were observed. Those output variations showed different trends depending on the dose-rate. The analysis of the relative output variation as a function of various beam parameters showed that in a given configuration, the dose-rate could be reliably varied over three orders of magnitude. Interdependence effects on the output variation between the parameters were also observed. The beam energy and field size were found to be slightly dose-rate-dependent and suitable mainly for small animal irradiation. The beam monitoring system was able to measure in a reproducible way the total charge of electrons that exit the machine, as long as the electron pulse amplitude remains above a given threshold. Furthermore, we were able to relate the charge measured with the monitoring system to the absorbed dose in a solid water phantom.
The Oriatron eRT6 was successfully commissioned for preclinical use and is currently in full operation, with studies being performed on the radiobiological effects of high dose-per-pulse irradiation.

Mots-clé
Commissioning, Linac, beam monitoring, high dose-per-pulse, high dose-rate, commissioning, linac
Pubmed
Web of science
Création de la notice
07/12/2017 19:31
Dernière modification de la notice
20/08/2019 16:18
Données d'usage