Pediatric CT: how to compare the Present DRL Values? : 12.10
Titre de la conférence
Swiss Radiological Congress 2008
Auteur⸱e⸱s
Verdun F., Gutierrez D., Alamo Maestre L., Bochud F., Gudinchet F.
Adresse
St. Gallen, Switzerland, May 29-31, 2008
ISBN
1424-4985
Statut éditorial
Publié
Date de publication
2008
Volume
8
Série
Swiss Medical Forum = Forum Médical Suisse
Pages
20
Langue
anglais
Résumé
Purpose: To determine dose thresholds, in term of CTDIvol, where subtle anatomical structures of pediatric CT images becomes no more detectable and compare them to the most recent Reference Dose Levels (DRL) proposed in the UK, Germany and Switzerland. Materials and methods: A GE LightSpeed-Ultra scanner (MSCT 8 slices) was used to perform chest and abdomen acquisitions on 8 patients (age range 2 to 16 years old) to provide a set of gold standard images. Dose reductions were then simulated by introducing image noise on raw data to provide simulated CT images with CTDIvol ranging from 2 to 22 mGy. All images were reviewed and scored independently by four experienced radiologists using the VGA methodology (Visual Grading Analysis) to determine the dose threshold where a significant loss of normal anatomy conspicuity appeared. Data were analyzed with ANOVA and Tukey HSD tests, a p >0.05 was considered to be significant. Results: No significant difference in VGA scoring appeared for CTDIvol leading to image noise levels lower than 10 and 25 HU for respectively abdominal and chest acquisitions. These data can thus be used to set the AEC (automatic exposure control) system of units having similar noise properties than the GE LightSpeed-Ultra used in this study. The present DRLs proposed for pediatric CT acquisitions are compatible with an excellent image quality level. Conclusion: The differences of DRL values proposed in Europe for pediatric acquisitions are marginal and assure a very good image quality level. The results of this study allow to further optimize the acquisition protocol by giving Noise Index value to set the AEC device.