Correcting Interdevice Bias of Horizontal White-to-White and Sulcus-to-Sulcus Measures Used for Implantable Collamer Lens Sizing.
Détails
Télécharger: BIB_7C8390100E47.P001.pdf (507.69 [Ko])
Etat: Public
Version: Author's accepted manuscript
Etat: Public
Version: Author's accepted manuscript
ID Serval
serval:BIB_9E7E9A5373A1
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Correcting Interdevice Bias of Horizontal White-to-White and Sulcus-to-Sulcus Measures Used for Implantable Collamer Lens Sizing.
Périodique
American Journal of Ophthalmology
ISSN
1879-1891 (Electronic)
ISSN-L
0002-9394
Statut éditorial
Publié
Date de publication
01/2016
Peer-reviewed
Oui
Volume
161
Pages
116-25.e1
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Publication Status: ppublish
Résumé
PURPOSE: To assess the agreement and repeatability of horizontal white-to-white (WTW) and horizontal sulcus-to-sulcus (STS) diameter measurements and use these data in combination with available literature to correct for interdevice bias in preoperative implantable collamer lens (ICL) size selection.
DESIGN: Interinstrument reliability and bias assessment study.
METHODS: A total of 107 eyes from 56 patients assessed for ICL implantation at our institution were included in the study. This was a consecutive series of all patients with suitable available data. The agreement and bias between WTW (measured with the Pentacam and BioGraph devices) and STS (measured with the HiScan device) were estimated.
RESULTS: The mean spherical equivalent was -8.93 ± 5.69 diopters. The BioGraph measures of WTW were wider than those taken with the Pentacam (bias = 0.26 mm, P < .01), and both horizontal WTW measures were wider than the horizontal STS measures (bias >0.91 mm, P < .01). The repeatability (Sr) of STS measured with the HiScan was 0.39 mm, which was significantly reduced (Sr = 0.15 mm) when the average of 2 measures was used. Agreement between the horizontal WTW measures and horizontal STS estimates when bias was accounted for was г = 0.54 with the Pentacam and г = 0.64 with the BioGraph.
CONCLUSIONS: Large interdevice bias was observed for WTW and STS measures. STS measures demonstrated poor repeatability, but the average of repeated measures significantly improved repeatability. In order to conform to the US Food and Drug Administration's accepted guidelines for ICL sizing, clinicians should be aware of and account for the inconsistencies between devices.
DESIGN: Interinstrument reliability and bias assessment study.
METHODS: A total of 107 eyes from 56 patients assessed for ICL implantation at our institution were included in the study. This was a consecutive series of all patients with suitable available data. The agreement and bias between WTW (measured with the Pentacam and BioGraph devices) and STS (measured with the HiScan device) were estimated.
RESULTS: The mean spherical equivalent was -8.93 ± 5.69 diopters. The BioGraph measures of WTW were wider than those taken with the Pentacam (bias = 0.26 mm, P < .01), and both horizontal WTW measures were wider than the horizontal STS measures (bias >0.91 mm, P < .01). The repeatability (Sr) of STS measured with the HiScan was 0.39 mm, which was significantly reduced (Sr = 0.15 mm) when the average of 2 measures was used. Agreement between the horizontal WTW measures and horizontal STS estimates when bias was accounted for was г = 0.54 with the Pentacam and г = 0.64 with the BioGraph.
CONCLUSIONS: Large interdevice bias was observed for WTW and STS measures. STS measures demonstrated poor repeatability, but the average of repeated measures significantly improved repeatability. In order to conform to the US Food and Drug Administration's accepted guidelines for ICL sizing, clinicians should be aware of and account for the inconsistencies between devices.
Mots-clé
Bias (Epidemiology), Biometry/methods, Corneal Topography, Humans, Lens Implantation, Intraocular, Limbus Corneae/anatomy & histology, Microscopy, Acoustic, Myopia, Degenerative/surgery, Nomograms, Phakic Intraocular Lenses, Reproducibility of Results, Sclera/anatomy & histology
Pubmed
Web of science
Création de la notice
02/02/2016 17:22
Dernière modification de la notice
20/08/2019 15:04