Balloon-assisted coil embolization and large stent delivery for cerebral aneurysms with a new generation of dual lumen balloons (Copernic 2L).

Détails

ID Serval
serval:BIB_9BD3E551B4CF
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Balloon-assisted coil embolization and large stent delivery for cerebral aneurysms with a new generation of dual lumen balloons (Copernic 2L).
Périodique
Journal of neurointerventional surgery
Auteur⸱e⸱s
Guenego A., Zerlauth J.B., Puccinelli F., Hajdu S., Rotzinger D.C., Zibold F., Piechowiak E.I., Mordasini P., Gralla J., Dobrocky T., Daniel R.T., Chapot R., Mosimann P.J.
ISSN
1759-8486 (Electronic)
ISSN-L
1759-8478
Statut éditorial
Publié
Date de publication
04/2018
Peer-reviewed
Oui
Volume
10
Numéro
4
Pages
395-400
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
Dual coaxial lumen balloon microcatheters through which small stents can be delivered have recently been described. We report a series of a new type of dual lumen balloon catheter with a parallel lumen design enabling enhanced inflation and deflation properties through which larger stents may be deployed, including flow diverters (FD).
All aneurysms that were treated with a Copernic 2L (COP2L) dual lumen balloon catheter at our institution between February 2014 and December 2016 were assessed. Patient demographics, aneurysm characteristics, clinical and angiographic follow-up, as well as adverse events were analyzed.
A total of 18 aneurysms in 16 patients (14 women) were treated with the COP2L. Mean maximal aneurysm diameter was 6.4 mm, mean neck size was 3.3 mm (min 1; max 6.3), and mean aneurysm height/width was 1.1 (min 0.5; max 2.1). The COP2L was used for balloon-remodeled coiling exclusively in 2 aneurysms; coiling and FD stenting in 8; coiling and braided stent delivery in 3; coiling, braided and FD stenting in 1; and FD stenting without coiling in 4 (stenting alone). The rate of Roy-Raymond 1 (complete occlusion) changed from 22% in the immediate postoperative period to 100% at 3 months (mean imaging follow-up 8.2 months). There were three technical complications (3/16, 18.7%), including a perforation and two thromboembolic asymptomatic events that were rapidly controlled with the COP2L. There was no immediate or delayed morbidity or mortality (modified Rankin Scale score 0-1 in 100% of patients).
The COP2L is a new type of dual lumen balloon catheter that may be useful for balloon and/or stent-assisted coiling of cerebral aneurysms. The same device can be used to deliver stents up to 4.5 mm and to optimize stent/wall apposition or serve as a life-saving tool in case of thromboembolic or hemorrhagic events. Long-term efficacy and safety need to be further assessed with larger case-controlled cohorts.