Tip malposition of peripherally inserted central catheters: a prospective randomized controlled trial to compare bedside insertion to fluoroscopically guided placement.

Détails

ID Serval
serval:BIB_9B3EF6914D2A
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Tip malposition of peripherally inserted central catheters: a prospective randomized controlled trial to compare bedside insertion to fluoroscopically guided placement.
Périodique
European radiology
Auteur⸱e⸱s
Glauser F., Breault S., Rigamonti F., Sotiriadis C., Jouannic A.M., Qanadli S.D.
ISSN
1432-1084 (Electronic)
ISSN-L
0938-7994
Statut éditorial
Publié
Date de publication
07/2017
Peer-reviewed
Oui
Volume
27
Numéro
7
Pages
2843-2849
Langue
anglais
Notes
Publication types: Journal Article ; Randomized Controlled Trial
Publication Status: ppublish
Résumé
Peripherally inserted central catheter (PICC) use continues to increase, leading to the development of a blind bedside technique (BST) for placement. The aim of our study was to compare the BST with the fluoroscopically guided technique (FGT), with specific regard to catheter tip position (CTP).
One hundred eighty patients were randomized to either the BST or the FGT. All procedures were done by the same interventional team and included postprocedural chest X-ray to assess CTP. Depending on the international guidelines for optimal CTP, patients were classified in three types: optimal, suboptimal not needing repositioning, and nonoptimal requiring additional repositioning procedures. Fisher's test was used for comparisons.
One hundred seventy-one PICCs were successful inserted. In the BST groups, 23.3% of placements were suboptimal and 30% nonoptimal, requiring repositioning. In the FGT group, 5.6% were suboptimal and 1.1% nonoptimal. Thus, suboptimal and nonoptimal CTP were significantly lower in the FGT group (p < 0.001).
Tip malposition rates are high when using blind BST, exposing the patient to an increased risk of deep venous thrombosis and catheter malfunction. Using the FGT or emerging technologies that could help tip positioning are recommended, especially for long-term indications.
• Bedside and fluoroscopy guided techniques are commonly used for PICC placement. • Catheter malposition is the major technical issue with the bedside technique. • Catheter malposition occurred in 53% of patients with the bedside technique.

Mots-clé
Catheterization, Central Venous/adverse effects, Catheterization, Central Venous/instrumentation, Catheterization, Peripheral/instrumentation, Central Venous Catheters/adverse effects, Equipment Failure, Female, Fluoroscopy/methods, Follow-Up Studies, Humans, Male, Middle Aged, Point-of-Care Testing, Prospective Studies, Venous Thrombosis/diagnosis, Venous Thrombosis/etiology, Blind bedside technique, Fluoroscopically guided technique, Intensive care unit, Peripherally inserted central catheters, Tip malposition
Pubmed
Web of science
Création de la notice
21/12/2016 16:49
Dernière modification de la notice
20/08/2019 15:02
Données d'usage