Image texture, low contrast liver lesion detectability and impact on dose: Deep learning algorithm compared to partial model-based iterative reconstruction.

Détails

ID Serval
serval:BIB_9B062A5F3328
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Image texture, low contrast liver lesion detectability and impact on dose: Deep learning algorithm compared to partial model-based iterative reconstruction.
Périodique
European journal of radiology
Auteur⸱e⸱s
Racine D., Brat H.G., Dufour B., Steity J.M., Hussenot M., Rizk B., Fournier D., Zanca F.
ISSN
1872-7727 (Electronic)
ISSN-L
0720-048X
Statut éditorial
Publié
Date de publication
08/2021
Peer-reviewed
Oui
Volume
141
Pages
109808
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
To compare deep learning (True Fidelity, TF) and partial model based Iterative Reconstruction (ASiR-V) algorithm for image texture, low contrast lesion detectability and potential dose reduction.
Anthropomorphic phantoms (mimicking non-overweight and overweight patient), containing lesions of 6 mm in diameter with 20 HU contrast, were scanned at five different dose levels (2,6,10,15,20 mGy) on a CT system, using clinical routine protocols for liver lesion detection. Images were reconstructed using ASiR-V 0% (surrogate for FBP), 60 % and TF at low, medium and high strength. Noise texture was characterized by computing a normalized Noise Power Spectrum filtered by an eye filter. The similarity against FBP texture was evaluated using peak frequency difference (PFD) and root mean square deviation (RMSD). Low contrast detectability was assessed using a channelized Hotelling observer and the area under the ROC curve (AUC) was used as figure of merit. Potential dose reduction was calculated to obtain the same AUC for TF and ASiR-V.
FBP-like noise texture was more preserved with TF (PFD from -0.043mm-1 to -0.09mm-1, RMSD from 0.12mm-1 to 0.21mm-1) than with ASiR-V (PFD equal to 0.12 mm-1, RMSD equal to 0.53mm-1), resulting in a sharper image. AUC was always higher with TF than ASIR-V. In average, TF compared to ASiR-V, enabled a radiation dose reduction potential of 7%, 25 % and 33 % for low, medium and high strength respectively.
Compared to ASIR-V, TF at high strength does not impact noise texture and maintains low contrast liver lesions detectability at significant lower dose.