A new equine pericardial stentless valve

Détails

ID Serval
serval:BIB_982946283227
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
A new equine pericardial stentless valve
Périodique
Journal of Thoracic and Cardiovascular Surgery
Auteur⸱e⸱s
Mueller  X. M., von Segesser  L. K.
ISSN
0022-5223
Statut éditorial
Publié
Date de publication
06/2003
Peer-reviewed
Oui
Volume
125
Numéro
6
Pages
1405-11
Notes
Comparative Study
Journal Article --- Old month value: Jun
Résumé
OBJECTIVE: Small aortic valve replacement remains a challenging hemodynamic problem. A new bioprosthesis (3F Therapeutics, Lake Forest, Calif) was designed to further improve the hemodynamic performance currently achieved with stentless bioprostheses. This valve consists of a tubular structure assembled from 3 equal sections of equine pericardial material, with virtually no foreign material except for a thin polyester ring. Its hemodynamic performance was compared with that of a commercially available stentless prosthesis in a bovine model. Patients and Methods: Twelve calves (55 +/- 2.8 kg) received a 19-mm 3F valve (3F group, n = 6) or a 19-mm stentless control valve (control group, n = 6). The animals were fully equipped for hemodynamic monitoring and transvalvular gradient measurements. After implantation, dopamine was infused in increasing doses, and the hemodynamic values were recorded at each step of 100-microg/min increase. Linear regression analysis was applied for group comparison of each variable. RESULTS: The mean transvalvular gradient at 4.5 L/min was 3.48 +/- 0.14 mm Hg (95% confidence interval) in the 3F group and 5.72 +/- 0.28 mm Hg in the control group (P <.0001) and at 6.5 L/min, 7.4 +/- 1.55 mm Hg, and 11.13 +/- 0.18 mm Hg, respectively (P <.0001). The effective orifice area at 4.5 L/min was 2.4 +/- 0.03 cm(2) in the 3F group and 1.86 +/- 0.02 cm(2) in the control group (P <.0001) and at 6.5 L/min, 2.41 +/- 0.04 cm(2), and 1.96 +/- 0.02 cm(2), respectively (P <.0001). CONCLUSIONS: This new bioprosthesis without a stent and without a supporting wall that has its commissures fixed directly to the aorta outperforms in vivo standard stentless prostheses in the immediate postimplant period.
Mots-clé
Animals Aortic Valve *Bioprosthesis Cattle *Heart Valve Prosthesis Hemodynamics Prosthesis Design
Pubmed
Web of science
Open Access
Oui
Création de la notice
14/02/2008 15:18
Dernière modification de la notice
20/08/2019 15:59
Données d'usage