Design and testing of a high-flow autoperfusion catheter: an experimental study.

Détails

ID Serval
serval:BIB_8A2AE01D35E3
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Titre
Design and testing of a high-flow autoperfusion catheter: an experimental study.
Périodique
Journal of Vascular and Interventional Radiology
Auteur⸱e⸱s
Williams D.M., Marx M.V., Andrews J.C., Doenz F., Abrams G.D.
ISSN
1051-0443
Statut éditorial
Publié
Date de publication
05/1992
Peer-reviewed
Oui
Volume
3
Numéro
2
Pages
285-290
Langue
anglais
Résumé
An autoperfusion catheter is similar to an angioplasty balloon catheter with side holes in the guide-wire lumen proximal to the balloon. When the balloon of the autoperfusion catheter is deployed and inflated in an artery, the guide wire is removed, and the hub of the guide-wire lumen is capped. The catheter then allows passive distal perfusion by using ambient pressure to drive blood into the guide-wire lumen, through the balloon, and out the end hole. This article discusses the requirements and constraints of a high-flow autoperfusion catheter, summarizes attempts to modify standard angioplasty catheters for use as an autoperfusion catheter, and describes the design and testing of a custom autoperfusion catheter capable of delivering approximately 3 mL/sec at physiologic pressures. In a model of canine acute renal artery occlusion lasting 90 minutes, the custom autoperfusion catheter provided marked protection from acute tubular necrosis compared with conventional percutaneous transluminal angioplasty catheters. The authors conclude that the high-flow autoperfusion catheter may be useful as a temporary stent in cases of rupture, dissection, or penetrating wounds involving large arteries.
Mots-clé
Angioplasty, Balloon/adverse effects, Angioplasty, Balloon/instrumentation, Animals, Balloon Dilatation/adverse effects, Balloon Dilatation/instrumentation, Dogs, Equipment Design, Kidney/injuries, Kidney/pathology, Renal Artery Obstruction/therapy, Renal Circulation, Rheology
Pubmed
Création de la notice
11/04/2008 11:46
Dernière modification de la notice
20/08/2019 14:49