Quality assessment of cardiovascular magnetic resonance in the setting of the European CMR registry: description and validation of standardized criteria.

Détails

Ressource 1Télécharger: BIB_7F5C95D0AE3C.P001.pdf (2476.09 [Ko])
Etat: Public
Version: de l'auteur⸱e
ID Serval
serval:BIB_7F5C95D0AE3C
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Quality assessment of cardiovascular magnetic resonance in the setting of the European CMR registry: description and validation of standardized criteria.
Périodique
Journal of Cardiovascular Magnetic Resonance
Auteur⸱e⸱s
Klinke V., Muzzarelli S., Lauriers N., Locca D., Vincenti G., Monney P., Lu C., Nothnagel D., Pilz G., Lombardi M., van Rossum A.C., Wagner A., Bruder O., Mahrholdt H., Schwitter J.
ISSN
1532-429X (Electronic)
ISSN-L
1097-6647
Statut éditorial
Publié
Date de publication
2013
Volume
15
Pages
55
Langue
anglais
Notes
Publication types: Journal Article ; Research Support, Non-U.S. Gov'tPublication Status: epublish
Résumé
BACKGROUND: Cardiovascular magnetic resonance (CMR) has become an important diagnostic imaging modality in cardiovascular medicine. However, insufficient image quality may compromise its diagnostic accuracy. We aimed to describe and validate standardized criteria to evaluate a) cine steady-state free precession (SSFP), b) late gadolinium enhancement (LGE), and c) stress first-pass perfusion images. These criteria will serve for quality assessment in the setting of the Euro-CMR registry.
METHODS: Thirty-five qualitative criteria were defined (scores 0-3) with lower scores indicating better image quality. In addition, quantitative parameters were measured yielding 2 additional quality criteria, i.e. signal-to-noise ratio (SNR) of non-infarcted myocardium (as a measure of correct signal nulling of healthy myocardium) for LGE and % signal increase during contrast medium first-pass for perfusion images. These qualitative and quantitative criteria were assessed in a total of 90 patients (60 patients scanned at our own institution at 1.5T (n=30) and 3T (n=30) and in 30 patients randomly chosen from the Euro-CMR registry examined at 1.5T). Analyses were performed by 2 SCMR level-3 experts, 1 trained study nurse, and 1 trained medical student.
RESULTS: The global quality score was 6.7±4.6 (n=90, mean of 4 observers, maximum possible score 64), range 6.4-6.9 (p=0.76 between observers). It ranged from 4.0-4.3 for 1.5T (p=0.96 between observers), from 5.9-6.9 for 3T (p=0.33 between observers), and from 8.6-10.3 for the Euro-CMR cases (p=0.40 between observers). The inter- (n=4) and intra-observer (n=2) agreement for the global quality score, i.e. the percentage of assignments to the same quality tertile ranged from 80% to 88% and from 90% to 98%, respectively. The agreement for the quantitative assessment for LGE images (scores 0-2 for SNR <2, 2-5, >5, respectively) ranged from 78-84% for the entire population, and 70-93% at 1.5T, 64-88% at 3T, and 72-90% for the Euro-CMR cases. The agreement for perfusion images (scores 0-2 for %SI increase >200%, 100%-200%,<100%, respectively) ranged from 81-91% for the entire population, and 76-100% at 1.5T, 67-96% at 3T, and 62-90% for the Euro-CMR registry cases. The intra-class correlation coefficient for the global quality score was 0.83.
CONCLUSIONS: The described criteria for the assessment of CMR image quality are robust with a good inter- and intra-observer agreement. Further research is needed to define the impact of image quality on the diagnostic and prognostic yield of CMR studies.
Pubmed
Web of science
Open Access
Oui
Création de la notice
11/08/2013 8:45
Dernière modification de la notice
20/08/2019 14:40
Données d'usage