Quality index for the quantification of the information recorded along standard microelectrode tracks to the subthalamic nucleus in parkinsonian patients.
Détails
ID Serval
serval:BIB_7D841A4F7613
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Quality index for the quantification of the information recorded along standard microelectrode tracks to the subthalamic nucleus in parkinsonian patients.
Périodique
Neurophysiologie Clinique = Clinical Neurophysiology
ISSN
0987-7053 (Print)
ISSN-L
0987-7053
Statut éditorial
Publié
Date de publication
2004
Volume
34
Numéro
5
Pages
209-215
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Publication Status: ppublish
Résumé
OBJECTIVE: To quantify the usefulness of the neuronal activity recorded on a standard microelectrode track to the subthalamic nucleus (STN) for the determination of the transition between the thalamus and the STN.
METHODS: The study is based on analysis of 689 extracelullar single units recorded on 70 tracks passing through the thalamus and the STN. Using four neuron parameters that were correlated with electrode depth, a quality index (QI) for each track was computed and compared with the subjective assessment by the electrophysiologist of the track quality.
RESULTS: Subjectively, the transition between the thalamus and the STN was detected in 49 tracks (usual track) and not detected on 21 tracks (unusual tracks). Objectively, spike frequency, cell burst index (BI), signal relative root mean square (RMS) and spike relative amplitude were correlated with electrode depth and used to compute track QI. The average QI index of usual and unusual tracks was 0.25 +/- 0.9 and 0.85 +/- 0.15 (mean +/- confidence interval at P < 0.001), respectively. In 20 patients, QI correlates with post-operative measurement of electrode length in the STN.
CONCLUSION: These results demonstrate that simple statistical analysis taking into account the variation of single-unit characteristics with electrode depth can discriminate between useful and useless tracks for the determination of the STN localisation.
METHODS: The study is based on analysis of 689 extracelullar single units recorded on 70 tracks passing through the thalamus and the STN. Using four neuron parameters that were correlated with electrode depth, a quality index (QI) for each track was computed and compared with the subjective assessment by the electrophysiologist of the track quality.
RESULTS: Subjectively, the transition between the thalamus and the STN was detected in 49 tracks (usual track) and not detected on 21 tracks (unusual tracks). Objectively, spike frequency, cell burst index (BI), signal relative root mean square (RMS) and spike relative amplitude were correlated with electrode depth and used to compute track QI. The average QI index of usual and unusual tracks was 0.25 +/- 0.9 and 0.85 +/- 0.15 (mean +/- confidence interval at P < 0.001), respectively. In 20 patients, QI correlates with post-operative measurement of electrode length in the STN.
CONCLUSION: These results demonstrate that simple statistical analysis taking into account the variation of single-unit characteristics with electrode depth can discriminate between useful and useless tracks for the determination of the STN localisation.
Mots-clé
Electrophysiology, Humans, Magnetic Resonance Imaging, Microelectrodes, Monitoring, Intraoperative, Neurons/physiology, Parkinson Disease/physiopathology, Reproducibility of Results, Subthalamic Nucleus/physiopathology, Thalamus/physiopathology
Pubmed
Web of science
Création de la notice
25/01/2008 12:49
Dernière modification de la notice
20/08/2019 14:38