Correlation Between Electroencephalography and Automated Pupillometry in Critically Ill Patients: A Pilot Study.

Détails

ID Serval
serval:BIB_7D1580213B94
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Correlation Between Electroencephalography and Automated Pupillometry in Critically Ill Patients: A Pilot Study.
Périodique
Journal of neurosurgical anesthesiology
Auteur⸱e⸱s
Hasan S., Peluso L., Ferlini L., Legros B., Calabrò L., Oddo M., Creteur J., Vincent J.L., Gaspard N., Taccone F.S.
ISSN
1537-1921 (Electronic)
ISSN-L
0898-4921
Statut éditorial
Publié
Date de publication
01/04/2021
Peer-reviewed
Oui
Volume
33
Numéro
2
Pages
161-166
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
Electroencephalography (EEG) is widely used in the monitoring of critically ill comatose patients, but its interpretation is not straightforward. The aim of this study was to evaluate whether there is a correlation between EEG background pattern/reactivity to stimuli and automated pupillometry in critically ill patients.
Prospective assessment of pupillary changes to light stimulation was obtained using an automated pupillometry (NeuroLight Algiscan, ID-MED, Marseille, France) in 60 adult patients monitored with continuous EEG. The degree of encephalopathy and EEG reactivity were scored by 3 independent neurophysiologists blinded to the patient's history. The median values of baseline pupil size, pupillary constriction, constriction velocity, and latency were collected for both eyes. To assess sensitivity and specificity, we calculated areas under the receiver-operating characteristic curve.
The degree of encephalopathy assessed by EEG was categorized as mild (42%), moderate (37%), severe (10%) or suppression-burst/suppression (12%); a total of 47/60 EEG recordings were classified as "reactive." There was a significant difference in pupillary size, constriction rate, and constriction velocity, but not latency, among the different EEG categories of encephalopathy. Similarly, reactive EEG tracings were associated with greater pupil size, pupillary constriction rate, and constriction velocity compared with nonreactive recordings; there were no significant differences in latency. Pupillary constriction rate values had an area under the curve of 0.83 to predict the presence of severe encephalopathy or suppression-burst/suppression, with a pupillary constriction rate of < 20% having a sensitivity of 85% and a specificity of 79%.
Automated pupillometry can contribute to the assessment of cerebral dysfunction in critically ill patients.