Feasibility and electromagnetic compatibility study of the ClearPEM front-end electronics for simultaneous PET-MR imaging

Détails

ID Serval
serval:BIB_7BF783CB563C
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Feasibility and electromagnetic compatibility study of the ClearPEM front-end electronics for simultaneous PET-MR imaging
Périodique
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Auteur⸱e⸱s
Neves J.A., Bugalho R., Gruetter R., Magill A. W., Ortigao C., Silva J.C., Silva R., Varela J.
ISSN
0168-9002
Statut éditorial
Publié
Date de publication
2013
Volume
702
Pages
34-36
Langue
anglais
Résumé
In this work we present a first feasibility study of the ClearPEM technology for simultaneous PET-MR imaging. The mutual electromagnetic interference (EMI) effects between both systems were evaluated on a 7 T magnet by characterizing the response behavior of the ClearPEM detectors and front-end electronics to pulsed RF power and switched magnetic field gradients; and by analyzing the MR system performance degradation from noise pickup into the RF receiver chain, and from magnetic susceptibility artifacts caused by PET front-end materials.
Mots-clé
PET, ClearPEM, 7 Tesla MR, Simultaneous PET-MRI, EMI and EMC