Unacceptable failure of the PI2® implant.

Détails

Ressource 1Demande d'une copie Sous embargo indéterminé.
Accès restreint UNIL
Etat: Public
Version: Final published version
Licence: Non spécifiée
ID Serval
serval:BIB_760112C0479F
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Unacceptable failure of the PI2® implant.
Périodique
The Journal of hand surgery, European volume
Auteur⸱e⸱s
van Aaken J., Holzer N., Wehrli L., Delaquaize F., Gonzalez I.A., Beaulieu J.Y.
ISSN
2043-6289 (Electronic)
ISSN-L
0266-7681
Statut éditorial
Publié
Date de publication
11/2016
Peer-reviewed
Oui
Volume
41
Numéro
9
Pages
917-922
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
The PI2® spacer is designed for treatment of trapeziometacarpal (TM) osteoarthritis. However, the shape of this implant has raised concerns about its stability. We retrospectively investigated 45 implants in 41 patients treated for trapeziometacarpal osteoarthritis in our hospital between 2004 and 2009 who underwent trapeziectomy and insertion of a PI2® spacer. Outcome parameters included revision rates and clinical outcomes correlated with implant position and scaphometacarpal distance, assessed using standard radiographs. A total of 12 implants (27%) were removed at a median time of 10 months (interquartile range (IQR), 7-22 months). These included five dislocations and one early infection. Additionally, a further six patients underwent revision due to persistent pain. Three of these had scapho-trapezoid osteoarthritis, two had developed subluxation of the implant, and one did not show any radiographic abnormalities. A review of patient records revealed that 33 implants remained in place at a median time of 29 months (IQR, 20-57). However, of those, only 21 implants (64%) in 17 patients were available for clinical evaluation at a median follow-up of 29 months (IQR, 19-62 months). No significant differences in clinical outcomes including functional results were observed between in-place ( n = 8) and subluxated ( n = 13) implants. Due to the high revision rate (12/45), consistent with other reports in the literature, we have abandoned the use of the PI2® spacer. We recommend the establishment of a registry for evaluation of future implants.
Case-series study/level IV.
Mots-clé
Aged, Aged, 80 and over, Carpometacarpal Joints, Female, Hand Strength, Humans, Joint Prosthesis/adverse effects, Male, Middle Aged, Osteoarthritis/surgery, Patient Satisfaction, Prosthesis Design, Prosthesis Failure, Retrospective Studies, Trapezium Bone/surgery, Treatment Outcome, PI2®, osteoarthritis, pyrocarbon, trapeziometacarpal joint
Pubmed
Web of science
Création de la notice
18/06/2016 17:56
Dernière modification de la notice
28/04/2020 5:20
Données d'usage