Valved stents for transapical pulmonary valve replacement.

Détails

ID Serval
serval:BIB_7404FD00A335
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Valved stents for transapical pulmonary valve replacement.
Périodique
Journal of Thoracic and Cardiovascular Surgery
Auteur⸱e⸱s
Huber C.H., Hurni M., Tsang V., von Segesser L.K.
ISSN
1097-685X
Statut éditorial
Publié
Date de publication
2009
Peer-reviewed
Oui
Volume
137
Numéro
4
Pages
914-918
Langue
anglais
Résumé
OBJECTIVES: Pulmonary valve insufficiency remains a leading cause for reoperations in congenital cardiac surgery. The current percutaneous approach is limited by the size of the access vessel and variable right ventricular outflow tract morphology. This study assesses the feasibility of transapical pulmonary valve replacement based on a new valved stent construction concept. METHODS: A new valved stent design was implanted off-pump under continuous intracardiac echocardiographic and fluoroscopic guidance into the native right ventricular outflow tract in 8 pigs (48.5 +/- 6.0 kg) through the right ventricular apex, and device function was studied by using invasive and noninvasive measures. RESULTS: Procedural success was 100% at the first attempt. Procedural time was 75 +/- 15 minutes. All devices were delivered at the target site with good acute valve function. No valved stents dislodged. No animal had significant regurgitation or paravalvular leaking on intracardiac echocardiographic analysis. All animals had a competent tricuspid valve and no signs of right ventricular dysfunction. The planimetric valve orifice was 2.85 +/- 0.32 cm(2). No damage to the pulmonary artery or structural defect of the valved stents was found at necropsy. CONCLUSIONS: This study confirms the feasibility of direct access valve replacement through the transapical procedure for replacement of the pulmonary valve, as well as validity of the new valved stent design concept. The transapical procedure is targeting a broader patient pool, including the very young and the adult patient. The device design might not be restricted to failing conduits only and could allow for implantation in a larger patient population, including those with native right ventricular outflow tract configurations.
Mots-clé
Feasibility Studies, Heart Valve Prosthesis, Heart Valve Prosthesis Implantation, Humans, Prosthesis Design, Pulmonary Valve, Pulmonary Valve Insufficiency, Stents
Pubmed
Web of science
Open Access
Oui
Création de la notice
29/05/2009 10:27
Dernière modification de la notice
20/08/2019 15:31
Données d'usage