Minimally-invasive fixation of distal extra-articular femur fractures with locking plates: limitations and failures.

Détails

ID Serval
serval:BIB_7339CF87B202
Type
Article: article d'un périodique ou d'un magazine.
Sous-type
Etude de cas (case report): rapporte une observation et la commente brièvement.
Collection
Publications
Titre
Minimally-invasive fixation of distal extra-articular femur fractures with locking plates: limitations and failures.
Périodique
Orthopaedics and Traumatology, Surgery and Research
Auteur⸱e⸱s
Ehlinger M., Adam P., Arlettaz Y., Moor B.K., DiMarco A., Brinkert D., Bonnomet F.
ISSN
1877-0568 (Electronic)
ISSN-L
1877-0568
Statut éditorial
Publié
Date de publication
2011
Volume
97
Numéro
6
Pages
668-674
Langue
anglais
Notes
Publication types: Case Reports ; Journal ArticlePublication Status: ppublish
Résumé
Minimally-invasive fixation using a locking plate and early motion is normal practice. However, technical errors and pitfalls are common. This surgery has a set of rules that encompass both the mechanics of the internal fixation system and the implantation itself. If these rules are not strictly followed, alignment defects and/or early failure of the fixation can occur. We analysed four cases of clinical failure that were encountered after minimally-invasive distal femoral extra-articular fixation with locking plates. The following rules must be followed with this technique: extra-articular fracture, minimally-invasive approach, long plate alternating between locking screw and empty hole (five holes on either side of fracture), bi-cortical screws, placement of locking screws near a complex fracture but away from a simple fracture. Osteoporotic bone, obesity that interferes with the instrumentation, articular fracture, horizontal fracture line and surgeon experience are all limitations of this minimally-invasive technique.
Mots-clé
Aged, Aged, 80 and over, Bone Plates, Female, Femoral Fractures/surgery, Fracture Fixation, Internal/instrumentation, Fracture Fixation, Internal/methods, Humans, Surgical Procedures, Minimally Invasive, Treatment Failure
Pubmed
Web of science
Open Access
Oui
Création de la notice
23/09/2014 19:43
Dernière modification de la notice
20/08/2019 15:31
Données d'usage