Reversed shoulder arthroplasty baseplate fixed according to the three major columns principle

Détails

ID Serval
serval:BIB_677DC5DAF520
Type
Actes de conférence (partie): contribution originale à la littérature scientifique, publiée à l'occasion de conférences scientifiques, dans un ouvrage de compte-rendu (proceedings), ou dans l'édition spéciale d'un journal reconnu (conference proceedings).
Sous-type
Abstract (résumé de présentation): article court qui reprend les éléments essentiels présentés à l'occasion d'une conférence scientifique dans un poster ou lors d'une intervention orale.
Collection
Publications
Institution
Titre
Reversed shoulder arthroplasty baseplate fixed according to the three major columns principle
Titre de la conférence
Annual meeting of the Swiss Society of Orthopaedics and Traumatology
Auteur⸱e⸱s
Djahangiri A., Kwiatkowski B., Farron A.
Adresse
St. Gallen - Switzerland, 30 June - 2 July 2010
ISBN
1424-7860
Statut éditorial
Publié
Date de publication
2010
Peer-reviewed
Oui
Volume
140
Série
Swiss Medical Weekly
Pages
27S
Langue
anglais
Notes
Meeting Abstract
Résumé
Introduction: Glenoid bone volume and bone quality can render the fixation of a reversed shoulder arthroplasty (RSA) basis plate hazardous. Cadaveric study at our institution has demonstrated that optimal baseplate fixation could be achieved with screws in three major columns. Our aim is to review our early rate of aseptic glenoid loosening in a series of baseplates fixed according to this principle.
Methods: Between 2005 and 2008, 48 consecutive RSA (Reversed Aequalis) were implanted in 48 patients with an average age of 74.4 years (range, 56 to 86 years). There were 37 women and 11 men. Twenty-seven primary RSAs were performed for cuff tear arthropathy, 3 after failed rotator cuff surgery, 6 for failed arthroplasties, 7 for acute fractures and 5 after failed ORIF. All baseplate fixations were done using a nonlocking posterior screw in the scapular spine, a nonlocking anterior screw in the glenoid body, a locking superior screw in the coracoid and a locking inferior screw in the pillar. All patients were reviewed with standardized radiographs. We reported the positions of the screws in relation to the scapular spine and the coracoid process in two different views. We defined screw positions as totally, partially or out of the target. Finally, we reported aseptic glenoid loosening which was defined as implant subsidence.
Results: Four patients were lost to follow-up. Thus 44 shoulders could be reviewed after a mean follow-up of 16 months (range, 9 to 32 months). Thirty-seven (84%) screws were either partially or totally in the spine. Thus, 7 (16%) scapular spine screws were out of the target. No coracoid screw was out of the target. At final follow-up control, we reported no glenoid loosening.
Conclusion: Early glenoid loosening occurred before the two years follow-up and is most of time related to technical problems and/or insufficient glenoid bone stock and bone quality. Our study demonstrate that baseplate fixation of a RSA according to the three columns principle is a reproducible technique and a valuable way to prevent early glenoid loosening.
Web of science
Création de la notice
14/10/2010 12:53
Dernière modification de la notice
20/08/2019 15:23
Données d'usage