Implementation and validation of a beam-current transformer on a medical pulsed electron beam LINAC for FLASH-RT beam monitoring.

Détails

Ressource 1Demande d'une copie Sous embargo indéterminé.
Accès restreint UNIL
Etat: Public
Version: de l'auteur⸱e
Licence: CC BY 4.0
ID Serval
serval:BIB_670F6488D9CD
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Implementation and validation of a beam-current transformer on a medical pulsed electron beam LINAC for FLASH-RT beam monitoring.
Périodique
Journal of applied clinical medical physics
Auteur⸱e⸱s
Oesterle R., Gonçalves Jorge P., Grilj V., Bourhis J., Vozenin M.C., Germond J.F., Bochud F., Bailat C., Moeckli R.
ISSN
1526-9914 (Electronic)
ISSN-L
1526-9914
Statut éditorial
Publié
Date de publication
11/2021
Peer-reviewed
Oui
Volume
22
Numéro
11
Pages
165-171
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
To implement and validate a beam current transformer as a passive monitoring device on a pulsed electron beam medical linear accelerator (LINAC) for ultra-high dose rate (UHDR) irradiations in the operational range of at least 3 Gy to improve dosimetric procedures currently in use for FLASH radiotherapy (FLASH-RT) studies.
Two beam current transformers (BCTs) were placed at the exit of a medical LINAC capable of UHDR irradiations. The BCTs were validated as monitoring devices by verifying beam parameters consistency between nominal values and measured values, determining the relationship between the charge measured and the absorbed dose, and checking the short- and long-term stability of the charge-absorbed dose ratio.
The beam parameters measured by the BCTs coincide with the nominal values. The charge-dose relationship was found to be linear and independent of pulse width and frequency. Short- and long-term stabilities were measured to be within acceptable limits.
The BCTs were implemented and validated on a pulsed electron beam medical LINAC, thus improving current dosimetric procedures and allowing for a more complete analysis of beam characteristics. BCTs were shown to be a valid method for beam monitoring for UHDR (and therefore FLASH) experiments.
Mots-clé
Clinical Protocols, Electrons, Humans, Particle Accelerators, Radiation Monitoring, Radiometry, Radiotherapy Dosage, FLASH RT, beam current monitoring, ultra-high dose rate
Pubmed
Web of science
Open Access
Oui
Création de la notice
12/10/2021 10:00
Dernière modification de la notice
07/12/2021 7:37
Données d'usage