Ultrashort echo time imaging of the lungs under high-frequency noninvasive ventilation: A new approach to lung imaging.

Détails

Ressource 1Télécharger: 31136048_BIB_65119F8236F7.pdf (1080.96 [Ko])
Etat: Public
Version: Final published version
Licence: CC BY-NC-ND 4.0
ID Serval
serval:BIB_65119F8236F7
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Ultrashort echo time imaging of the lungs under high-frequency noninvasive ventilation: A new approach to lung imaging.
Périodique
Journal of magnetic resonance imaging
Auteur⸱e⸱s
Delacoste J., Dournes G., Dunet V., Ogna A., Noirez L., Simons J., Long O., Berchier G., Stuber M., Lovis A., Beigelman-Aubry C.
ISSN
1522-2586 (Electronic)
ISSN-L
1053-1807
Statut éditorial
Publié
Date de publication
12/2019
Peer-reviewed
Oui
Volume
50
Numéro
6
Pages
1789-1797
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
Although ultrashort echo time (UTE) sequences allow excellent assessment of lung parenchyma, image quality remains lower than that of computed tomography (CT).
To investigate a high-frequency noninvasive ventilation (HF-NIV) technique allowing a stabilized inspiration and to compare image quality with current dedicated MR sequences.
Prospective.
Ten healthy volunteers.
3D radial UTE sequence at 1.5T.
UTE-HF-NIV sequence was compared with UTE-free-breathing (UTE-FB), reconstructed at end expiration (UTE-Exp) and average (UTE-Avg), and breath-hold VIBE sequences. The distance from lung apex to the dome of the right hemidiaphragm was measured. Visual assessment of the visibility and sharpness of normal anatomical structures was carried out. Dedicated software also quantitatively evaluated vessel-lung and right lung-liver interface sharpness. Apparent signal ratio (Sr) and contrast ratios (Cr) were quantitatively evaluated.
Wilcoxon signed rank test for visual scores, paired t-test for continuous variables, significance at P < 0.05.
The distance between apex and the right hemidiaphragmatic dome was significantly larger (P < 0.001) with UTE-HF-NIV compared with UTE-FB and VIBE acquisitions. Vessel and airway visibility had identical median visual scores with all UTE methods. Median visual scores for sharpness of vessels and airways were significantly higher (P < 0.001) with HF-NIV (vessels = 3; airways = 2) than in UTE-FB (vessels = 2; airways = 1) and VIBE (vessels = 1; airways = 1). Software-based vessel sharpness evaluation resulted in larger values in 8/10 volunteers with UTE-HF-NIV (67.3 ± 9.8) compared with UTE-Avg (62.3 ± 12.6) but the average difference was not significant (P = 0.28). The sharpness of the lung-liver interface was significantly higher (P < 0.001) with HF-NIV (17.3 ± 5.3) compared with UTE-Avg (14.1 ± 3.9). Significantly higher values (P < 0.01) of Sr and Cr were observed with UTE-HF-NIV compared with UTE-FB and VIBE.
HF-NIV allowing acquisition at full inspiration significantly improves image quality for lung imaging. This could offer the option to alternate some follow-up CT studies by using this technique.
2 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2019;50:1789-1797.