Harmonization of LC-MS/MS Measurements of Plasma Free Normetanephrine, Metanephrine, and 3-Methoxytyramine.

Détails

ID Serval
serval:BIB_6314F9E194B8
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Harmonization of LC-MS/MS Measurements of Plasma Free Normetanephrine, Metanephrine, and 3-Methoxytyramine.
Périodique
Clinical chemistry
Auteur⸱e⸱s
Peitzsch M., Novos T., Kaden D., Kurlbaum M., van Herwaarden A.E., Müller D., Adaway J., Grouzmann E., McWhinney B., Hoad K., Woollard G., Kema I., Boot C., Fassnacht M., Sweep F., Loh T.P., Horvath A.R., Eisenhofer G.
ISSN
1530-8561 (Electronic)
ISSN-L
0009-9147
Statut éditorial
Publié
Date de publication
05/08/2021
Peer-reviewed
Oui
Volume
67
Numéro
8
Pages
1098-1112
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
Plasma-free normetanephrine and metanephrine (metanephrines) are the recommended biomarkers for testing of pheochromocytoma and paraganglioma (PPGL). This study evaluated the status of harmonization of liquid chromatography-tandem mass spectrometry-based measurements of plasma metanephrines and methoxytyramine and clinical interpretation of test results.
125 plasma samples from patients tested for PPGLs were analyzed in 12 laboratories. Analytical performance was also assessed from results of a proficiency-testing program. Agreement of test results from different laboratories was assessed by Passing-Bablok regression and Bland-Altman analysis. Agreement in clinical test interpretation based on laboratory specific reference intervals was also examined.
Comparisons of analytical test results by regression analysis revealed strong correlations for normetanephrine and metanephrine (R ≥ 0.95) with mean slopes of 1.013 (range 0.975-1.078), and 1.019 (range 0.963-1.081), and intercepts of -0.584 (-53.736 to 54.790) and -3.194 (-17.152 to 5.933), respectively. The mean bias between methods was 1.2% (-11.6% to 16.0%) for metanephrine and 0.1% (-18.0% to 9.5%) for normetanephrine. Measurements of 3-methoxytyramine revealed suboptimal agreement between laboratories with biases ranging from -32.2% to 64.0%. Interrater agreement in test interpretation was >94% for metanephrine and >84% for normetanephrine; improvements in interrater agreement were observed with use of harmonized reference intervals, including age-specific cut-offs for normetanephrine.
Analytical methods for metanephrines are well harmonized between laboratories. However, the 16% disagreement in test interpretation for normetanephrine suggests use of suboptimal method-dependent reference intervals for clinical decision-making for this metabolite. Improved analytical methods and reference interval harmonization are particularly required for 3-methoxytyramine.
Mots-clé
LC-MS/MS, harmonization, metanephrines, paraganglioma, pheochromocytoma, standardization