Validation of Monte Carlo dose calculation algorithm for CyberKnife multileaf collimator.

Détails

Ressource 1Télécharger: J Applied Clin Med Phys - 2021 - Gondr - Vali1.pdf (474.00 [Ko])
Etat: Public
Version: Final published version
Licence: CC BY 4.0
ID Serval
serval:BIB_5E1F27A9CF4B
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Validation of Monte Carlo dose calculation algorithm for CyberKnife multileaf collimator.
Périodique
Journal of applied clinical medical physics
Auteur⸱e⸱s
Gondré M., Marsolat F., Bourhis J., Bochud F., Moeckli R.
ISSN
1526-9914 (Electronic)
ISSN-L
1526-9914
Statut éditorial
Publié
Date de publication
02/2022
Peer-reviewed
Oui
Volume
23
Numéro
2
Pages
e13481
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
To commission and evaluate the Monte Carlo (MC) dose calculation algorithm for the CyberKnife equipped with a multileaf collimator (MLC).
We created a MC model for the MLC using an integrated module of the CyberKnife treatment planning software (TPS). Two parameters could be optimized: the maximum energy and the source full width at half-maximum (FWHM). The optimization was performed by minimizing the differences between the measured and the MC calculated tissue phantom ratios and profiles. MLC plans were calculated in the TPS with the MC algorithm and irradiated on different phantoms. The dose was measured using an A1SL ionization chamber and EBT3 Gafchromic films, and then compared to the TPS dose to obtain dose differences (ΔD). Finally, patient-specific quality assurances (QA) were performed with global gamma index criteria of 3%/1 mm.
The maximum energy and source FWHM showing the best agreement with measurements were 6.4 MeV and 1.8 mm. The output factors calculated with these parameters gave an agreement within ±1% with measurements. The ΔD showed that MC model systematically underestimated the dose with an average of -1.5% over all configurations tested. For depths deeper than 12 cm, the ΔD increased, up to -3.0% (maximum at 15.5 cm depth).
The MC model for MLC of CyberKnife is clinically acceptable but underestimates the delivered dose by an average of -1.5%. Therefore, we recommend using the MC algorithm with the MLC only in heterogeneous regions and for shallow-seated tumors.
Mots-clé
Algorithms, Humans, Monte Carlo Method, Phantoms, Imaging, Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted, CyberKnife, Monte Carlo dose algorithm, commissioning, multileaf collimator
Pubmed
Web of science
Open Access
Oui
Création de la notice
03/12/2021 10:26
Dernière modification de la notice
13/08/2022 7:10
Données d'usage