Evaluation of a new automated third-generation anti-HCV enzyme immunoassay.

Détails

ID Serval
serval:BIB_5521
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Evaluation of a new automated third-generation anti-HCV enzyme immunoassay.
Périodique
Journal of Clinical Laboratory Analysis
Auteur⸱e⸱s
Lavanchy D., Steinmann J., Möritz A., Frei P.C.
ISSN
0887-8013 (Print)
ISSN-L
0887-8013
Statut éditorial
Publié
Date de publication
1996
Volume
10
Numéro
5
Pages
269-276
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
The new Cobas Core Anti-HCV EIA was evaluated in two centers for its ability to detect antibodies directed to hepatitis C virus in human serum. This assay, which can be run fully automated on a random access analyzer, was compared with three other commercially available screening tests: the Ortho HCV 3.0 ELISA, the Murex anti-HCV, and the Abbott HCV EIA second generation. Positive or discrepant results were further investigated using the Wellcozyme HCV Western Blot or the Abbott Matrix HCV assays. The results obtained from analyzing 5045 serum samples showed a high correlation between the Cobas Core Anti-HCV EIA and the other screening assays, ranging from 98.9% to 99.9%. Diagnostic specificities and sensitivities ranged from 99.7% to 100% and from 98.8% to 100%, respectively. In this study, the Cobas Core Anti-HCV EIA proved to be a very convenient test, able to perform at the highest levels of sensitivity and specificity.
Mots-clé
Hepatitis C Antibodies/blood, Humans, Immunoenzyme Techniques, Sensitivity and Specificity
Pubmed
Web of science
Création de la notice
19/11/2007 13:42
Dernière modification de la notice
20/08/2019 15:09
Données d'usage