Unexpected wear of an unicompartimental knee arthroplasty in oxidized zirconium.
Détails
ID Serval
serval:BIB_54EA3379B554
Type
Article: article d'un périodique ou d'un magazine.
Sous-type
Etude de cas (case report): rapporte une observation et la commente brièvement.
Collection
Publications
Institution
Titre
Unexpected wear of an unicompartimental knee arthroplasty in oxidized zirconium.
Périodique
Acta orthopaedica Belgica
ISSN
0001-6462 (Print)
ISSN-L
0001-6462
Statut éditorial
Publié
Date de publication
12/2015
Volume
81
Numéro
4
Pages
790-795
Langue
anglais
Notes
Publication types: Case Reports ; Journal Article
Publication Status: ppublish
Publication Status: ppublish
Résumé
Unicompartimental knee arthroplasty is a successful procedure for the treatment of localized osteoarthritis to one compartment of the knee with good long-term results. However, several modes of failure of unicompartimental knee arthroplasty have been described, namely aseptic or septic loosening, progression of disease, wear, and instability. Metallosis after unicompartimental knee arthroplasty is rarely reported and is most often related with polyethylene wear or break. We report on a case of rapid failure of unicompartimental knee arthroplasty in oxidized zirconium associated with metallosis secondary to the dislocation of the polyethylene.
Mots-clé
Aged, Arthroplasty, Replacement, Knee/methods, Female, Humans, Knee Prosthesis, Metal-on-Metal Joint Prostheses, Osteoarthritis, Knee/surgery, Prosthesis Design, Zirconium