Use of accuracy profile for the validation of a gas chromatography-negative chemical ionization tandem mass spectrometry method: quantification of ethyl glucuronide in hair

Détails

ID Serval
serval:BIB_3EBEAC18A4C3
Type
Actes de conférence (partie): contribution originale à la littérature scientifique, publiée à l'occasion de conférences scientifiques, dans un ouvrage de compte-rendu (proceedings), ou dans l'édition spéciale d'un journal reconnu (conference proceedings).
Sous-type
Abstract (résumé de présentation): article court qui reprend les éléments essentiels présentés à l'occasion d'une conférence scientifique dans un poster ou lors d'une intervention orale.
Collection
Publications
Institution
Titre
Use of accuracy profile for the validation of a gas chromatography-negative chemical ionization tandem mass spectrometry method: quantification of ethyl glucuronide in hair
Titre de la conférence
Annales de toxicologie analytique
Auteur⸱e⸱s
Kharbouche H., Sporkert F., Augsburger M., Mangin P., Staub C.
Editeur
EDP Sciences S.A.
Organisation
Société Française de Toxicologie Analytique
ISBN
0768-598X
Statut éditorial
Publié
Date de publication
08/2009
Editeur⸱rice scientifique
Goullé J.P.
Volume
21
Série
Oral présentations : 076
Pages
S35
Langue
anglais
Résumé
Introduction: Ethylglucuronide (EtG) is a direct and specific metabolite of ethanol. Its determination in hair is of increasing interest for detecting and monitoring alcohol abuse. The quantification of EtG in hair requires analytical methods showing highest sensitivity and specificity. We present a fully validated method based on gas chromatography-negative chemical ionization tandem mass spectrometry (GC-NCI-MS/MS). The method was validated using French Society of Pharmaceutical Sciences and Techniques (SFSTP) guidelines which are based on the determination of the total measurement error and accuracy profiles.
Methods: Washed and powdered hair is extracted in water using an ultrasonic incubation. After purification by Oasis MAX solid phase extraction, the derivatized EtG is detected and quantified by GC-NCI-MS/MS method in the selected reaction monitoring mode. The transitions m/z 347 / 163 and m/z 347 / 119 were used for the quantification and identification of EtG. Four quality controls (QC) prepared with hair samples taken post mortem from 2 subjects with a known history of alcoholism were used. A proficiency test with 7 participating laboratories was first run to validate the EtG concentration of each QC sample. Considering the results of this test, these samples were then used as internal controls for validation of the method.
Results: The mean EtG concentrations measured in the 4 QC were 259.4, 130.4, 40.8, and 8.4 pg/mg hair. Method validation has shown linearity between 8.4 and 259.4 pg/mg hair (r2 > 0.999). The lower limit of quantification was set up at 8.4 pg/mg. Repeatability and intermediate precision were found less than 13.2% for all concentrations tested.
Conclusion: The method proved to be suitable for routine analysis of EtG in hair. GC-NCI-MS/MS method was then successfully applied to the analysis of EtG in hair samples collected from different alcohol consumers.
Mots-clé
ethyl glucuronide, hair, accuracy profile, alcohol markers