An image quality comparison of standard and dual-side read CR systems for pediatric radiology.

Détails

ID Serval
serval:BIB_3D81E5BD7ADF
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
An image quality comparison of standard and dual-side read CR systems for pediatric radiology.
Périodique
Medical physics
Auteur⸱e⸱s
Monnin P., Holzer Z., Wolf R., Neitzel U., Vock P., Gudinchet F., Verdun F.R.
ISSN
0094-2405
Statut éditorial
Publié
Date de publication
2006
Peer-reviewed
Oui
Volume
33
Numéro
2
Pages
411-20
Langue
anglais
Notes
Publication types: Comparative Study ; Journal Article - Publication Status: ppublish
Résumé
An objective analysis of image quality parameters was performed for a computed radiography (CR) system using both standard single-side and prototype dual-side read plates. The pre-sampled modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) for the systems were determined at three different beam qualities representative of pediatric chest radiography, at an entrance detector air kerma of 5 microGy. The NPS and DQE measurements were realized under clinically relevant x-ray spectra for pediatric radiology, including x-ray scatter radiations. Compared to the standard single-side read system, the MTF for the dual-side read system is reduced, but this is offset by a significant decrease in image noise, resulting in a marked increase in DQE (+40%) in the low spatial frequency range. Thus, for the same image quality, the new technology permits the CR system to be used at a reduced dose level.
Mots-clé
Child, Equipment Failure Analysis, Humans, Phantoms, Imaging, Radiographic Image Enhancement, Radiographic Image Interpretation, Computer-Assisted, Radiometry, Reproducibility of Results, Scattering, Radiation, Sensitivity and Specificity, Spectrometry, X-Ray Emission