Update of national diagnostic reference levels for adult CT in Switzerland and assessment of radiation dose reduction since 2010.

Détails

ID Serval
serval:BIB_39CBA0F2E6C2
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Update of national diagnostic reference levels for adult CT in Switzerland and assessment of radiation dose reduction since 2010.
Périodique
European radiology
Auteur⸱e⸱s
Aberle C., Ryckx N., Treier R., Schindera S.
ISSN
1432-1084 (Electronic)
ISSN-L
0938-7994
Statut éditorial
Publié
Date de publication
03/2020
Peer-reviewed
Oui
Volume
30
Numéro
3
Pages
1690-1700
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
To update the national diagnostic reference levels (DRLs) for adult CT in Switzerland using dose management software and to compare them to the previous Swiss DRLs from 2010.
CT dose data from 14 radiological institutes with a total of 50 CT scanners were collected with locally installed dose management software between 2014 and 2017. Data were assigned to 15 defined master protocols. Data cleaning steps were developed and adjusted individually for each participating institute and protocol. The DRLs for each master protocol were calculated as the 75th percentile of the distribution of the median volume computed tomography dose index (CTDI <sub>vol</sub> ) and dose-length product (DLP) values per CT scanner.
In total, 220,269 CT exams were available after data cleaning. Updated DRLs showed a clear trend towards lower doses compared with previous DRLs. The average relative change in the DRLs for CTDI <sub>vol</sub> was - 30% (0 to - 47%) and - 22% for DLP (+ 20 to - 40%). The largest relative decrease in the DRL for DLP was observed for the cervical spine protocol (- 40%), the two chest protocols (chest, - 37%; and exclusion of pulmonary embolism, - 33%), and the two neck protocols (neck, - 32%; and carotid angiography, - 28%). The DRLs for other protocols, for example the head and the abdomen-pelvis protocol, showed smaller relative changes (- 11% and - 17%).
The updated national DRLs are substantially lower than the previous values from 2010, demonstrating technological progress and the efforts of the radiological community to lower CT radiation exposure.
• Dose management software allows the establishment of DRLs based on big data. • Updated Swiss DRLs for adult CT are substantially lower compared with those from 2010. • Swiss DRLs are low compared with other national DRLs.
Mots-clé
Radiation monitoring, Radiation protection, Switzerland, Tomography, X-ray computed, Tomography, X-ray computed
Pubmed
Web of science
Création de la notice
30/11/2019 13:37
Dernière modification de la notice
01/04/2023 6:51
Données d'usage