Field mapping without reference scan using asymmetric echo-planar techniques.

Détails

ID Serval
serval:BIB_39046B3DB0A3
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Titre
Field mapping without reference scan using asymmetric echo-planar techniques.
Périodique
Magnetic Resonance in Medicine
Auteur⸱e⸱s
Gruetter R., Tkác I.
ISSN
0740-3194 (Print)
ISSN-L
0740-3194
Statut éditorial
Publié
Date de publication
2000
Volume
43
Numéro
2
Pages
319-323
Langue
anglais
Notes
Publication types: Comparative Study ; Journal Article ; Research Support, U.S. Gov't, P.H.S.Publication Status: ppublish
Résumé
Improvements in Bo mapping and shimming were achieved by measuring the static field information in multiple subsequent echoes generated by an asymmetric echo-planar readout gradient train. With careful compensation, eddy current effects were shown to affect the adjustment of the shim coils minimally. In addition to reducing the time required for field mapping by two-fold, the sensitivity was simultaneously optimized irrespective of the prevalent T2* present, thereby minimizing the error of the static field measurement to below 0.1 Hz. With adiabatic low flip-angle excitation, the time required for field mapping was below 1 second.
Mots-clé
Echo-Planar Imaging/instrumentation, Echo-Planar Imaging/methods, Fourier Analysis, Humans, Phantoms, Imaging/statistics & numerical data, Software, Time Factors