F4H5: a novel substance for the removal of silicone oil from intraocular lenses.

Détails

ID Serval
serval:BIB_37F8F2FFA280
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Titre
F4H5: a novel substance for the removal of silicone oil from intraocular lenses.
Périodique
The British journal of ophthalmology
Auteur⸱e⸱s
Stappler T., Williams R., Wong D.
ISSN
1468-2079 (Electronic)
ISSN-L
0007-1161
Statut éditorial
Publié
Date de publication
03/2010
Peer-reviewed
Oui
Volume
94
Numéro
3
Pages
364-367
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
Adherent silicone oil on intraocular lenses (IOLs) following retinal detachment surgery induces large and irregular refractive errors and multiple images, and gives rise to glare, distorted and often poor vision. Its removal remains challenging, often requiring mechanical wiping or explantation. F4H5 is a new semifluorinated alkane into which silicone oil is readily soluble. The aim is to establish the effectiveness of F4H5 in removing silicone oil from three different types of IOL in vitro.
Silicone lenses (Tecnis ZM900, Advanced Medical Optics, Inc.), hydrophobic acrylic lenses (MA60, Alcon Laboratories, Inc.) and PMMA lenses (Ocular Vision, Inc) were first immersed in phosphate-buffered saline, second in silicone oil, then in F4H5 (Fluoron GmbH) for 10 min and lastly vigorously agitated in F4H5 for 1 min. They were weighed at each stage using scales accurate to 0.0001 g to measure the weight of the adherent oil. Dynamic contact angle (DCA) analysis was used to assess their surface properties.
Immersion in F4H5 alone removed 96.1% (+/-1.23) by weight of silicone oil from the hydrophobic acrylic lenses, 91.4% (+/-1.58) from the silicone and 95.6% (+/-1.44) from the PMMA IOLs. Immersion combined with 1 min of agitation increased the removal to 98.8% (+/-0.46) from the acrylic IOLs, to 93.7% (+/-0.48) from the silicone IOLs and to 100% (within +/-0.0001 g) from every PMMA IOL. After treatment with F4H5, all IOL were optically clear. DCA hysteresis curves remained permanently altered. All measurements were highly reproducible.
F4H5 was highly effective at removing the bulk of the silicone oil from all three groups of IOL. The DCA measurements suggested that their surface properties were permanently modified.
Mots-clé
Adhesiveness/drug effects, Drug Evaluation, Preclinical/methods, Fluorocarbons/pharmacology, Humans, Lenses, Intraocular, Silicone Oils, Solvents/pharmacology, Surface Properties/drug effects
Pubmed
Web of science
Création de la notice
30/08/2018 16:34
Dernière modification de la notice
20/08/2019 14:26
Données d'usage