Preclinical evaluation of implantable cardioverter-defibrillator developed for magnetic resonance imaging use.

Détails

Ressource 1Demande d'une copie Sous embargo indéterminé.
Accès restreint UNIL
Etat: Public
Version: de l'auteur⸱e
ID Serval
serval:BIB_34C68D276F85
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Preclinical evaluation of implantable cardioverter-defibrillator developed for magnetic resonance imaging use.
Périodique
Heart Rhythm
Auteur⸱e⸱s
Gold M.R., Kanal E., Schwitter J., Sommer T., Yoon H., Ellingson M., Landborg L., Bratten T.
ISSN
1556-3871 (Electronic)
ISSN-L
1547-5271
Statut éditorial
Publié
Date de publication
2015
Peer-reviewed
Oui
Volume
12
Numéro
3
Pages
631-638
Langue
anglais
Notes
Publication types: Journal Article Publication Status: ppublish
Résumé
BACKGROUND: Many patients with an implantable cardioverter-defibrillator (ICD) have indications for magnetic resonance imaging (MRI). However, MRI is generally contraindicated in ICD patients because of potential risks from hazardous interactions between the MRI and ICD system.
OBJECTIVE: The purpose of this study was to use preclinical computer modeling, animal studies, and bench and scanner testing to demonstrate the safety of an ICD system developed for 1.5-T whole-body MRI.
METHODS: MRI hazards were assessed and mitigated using multiple approaches: design decisions to increase safety and reliability, modeling and simulation to quantify clinical MRI exposure levels, animal studies to quantify the physiologic effects of MRI exposure, and bench testing to evaluate safety margin.
RESULTS: Modeling estimated the incidence of a chronic change in pacing capture threshold >0.5V and 1.0V to be less than 1 in 160,000 and less than 1 in 1,000,000 cases, respectively. Modeling also estimated the incidence of unintended cardiac stimulation to occur in less than 1 in 1,000,000 cases. Animal studies demonstrated no delay in ventricular fibrillation detection and no reduction in ventricular fibrillation amplitude at clinical MRI exposure levels, even with multiple exposures. Bench and scanner testing demonstrated performance and safety against all other MRI-induced hazards.
CONCLUSION: A preclinical strategy that includes comprehensive computer modeling, animal studies, and bench and scanner testing predicts that an ICD system developed for the magnetic resonance environment is safe and poses very low risks when exposed to 1.5-T normal operating mode whole-body MRI.