Evaluation of the Sysmex XQ-320 three-part differential haematology analyser and its flagging capabilities.

Détails

Ressource 1Télécharger: 38396348_BIB_31CC15F1E63D.pdf (1265.07 [Ko])
Etat: Public
Version: Final published version
Licence: CC BY 4.0
ID Serval
serval:BIB_31CC15F1E63D
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
Evaluation of the Sysmex XQ-320 three-part differential haematology analyser and its flagging capabilities.
Périodique
Journal of clinical laboratory analysis
Auteur⸱e⸱s
Coutaz C., Mamin A., Mintzas K.
ISSN
1098-2825 (Electronic)
ISSN-L
0887-8013
Statut éditorial
Publié
Date de publication
02/2024
Peer-reviewed
Oui
Volume
38
Numéro
4
Pages
e25017
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
Three-part differential (3PD) haematology analysers offer a quick, easy-to-use and economical way to acquire important information about a patient's physiology. In this study, we evaluated a new 3PD analyser, the Sysmex XQ-320, investigated its comparability with its predecessor (Sysmex XP-300) and the five-part differential analyser Sysmex XN-9000, and explored its flagging potential.
Analytical performance studies were conducted for repeatability, within-laboratory precision, between-day precision, carry-over and linearity with fresh blood and QC material. Method comparison was performed in 493 samples comparing XQ-320 with XP-300, using the XN-9000 as the gold standard.
The XQ-320 excelled manufacturer's specifications in the analytical performance studies, except for MXD in within-laboratory and between-day precisions using the QC material level 1. The XQ-320 showed correlation values greater than 0.94 with XN-9000 for the majority of the 20 reportable parameters (MXD# 0.891, MXD% 0.898 and MCHC 0.849). Improvements over the XP-300 were observed in WBC in the leucocytopenic range (bias -0.038 vs. -0.097) and PLT (bias 2.568 vs. -7.877, intercept 3.880 vs. -8.845). Concordance between XQ-320 and XP-300 was 91.9% for the WBC histogram abnormal distribution flag and 95.3% for the PLT flag. Patterns of increased neutrophils and decreased mixed cells on the XQ-320 were observed in samples that raised a flag on XN-9000.
The XQ-320 showed excellent analytical performance, and very good to excellent correlation with XN-9000 with improvements over XP-300. Flagging combined with parameter patterns identified additional suspected abnormal samples, thus making the XQ-320 an excellent solution for laboratories utilising 3PD analysers.
Mots-clé
Humans, Hematology, Laboratories, Nonoxynol, Blood Cell Count/methods, Reproducibility of Results, analytical performance, correlation, flagging, haematology analyser, three-part differential
Pubmed
Web of science
Open Access
Oui
Création de la notice
01/03/2024 13:11
Dernière modification de la notice
09/08/2024 14:57
Données d'usage