3-D Freehand Ultrasound Calibration Using a Tissue-Mimicking Phantom With Parallel Wires.

Détails

ID Serval
serval:BIB_288416C9B841
Type
Article: article d'un périodique ou d'un magazine.
Collection
Publications
Institution
Titre
3-D Freehand Ultrasound Calibration Using a Tissue-Mimicking Phantom With Parallel Wires.
Périodique
Ultrasound in medicine & biology
Auteur⸱e⸱s
Marcadent S., Heches J., Favre J., Desseauve D., Thiran J.P.
ISSN
1879-291X (Electronic)
ISSN-L
0301-5629
Statut éditorial
Publié
Date de publication
01/2023
Peer-reviewed
Oui
Volume
49
Numéro
1
Pages
165-177
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
This article describes a method used to calibrate 3-D freehand ultrasound systems based on phantoms with parallel wires forming two perpendicular planes, such as the usual general-purpose commercial phantoms. In our algorithm, the phantom pose is co-optimized with the calibration to avoid the need to precisely track the phantom. We provide a geometrical analysis to explain the proposed acquisition protocol. Finally, we give an estimate of the system accuracy and precision based on measurements acquired on an independent test phantom. We obtained error norms of 1.6 mm up to 6 cm of depth and 3.5 mm between 6 and 14 cm of depth, in total average. In conclusion, it is possible to calibrate ultrasound tracked-probe systems with a reasonable accuracy based on a general-purpose phantom. Contrarily to most calibration methods that imply the construction of the phantom, the present algorithm is based on a standard phantom geometry that is commercially available.
Mots-clé
Calibration, Imaging, Three-Dimensional/methods, Phantoms, Imaging, Ultrasonography/methods, Algorithms, Freehand 3-D ultrasound, Ultrasound tracked-probe system